Revision 4
122
12/15/2018
Compressibility Factor (Z)
The Compressibility Factor is calculated by equation 13-2.
Z = 1 – (P
b
×
Σ
Pn=1
[Conc
n
× b
n
])
13-2
Where:
Z
= Compressibility factor
P
b
= Base pressure (psia)
Conc
n
= Normalized concentration of component n
b
n
= ‘Summation factors’, as defined in GPA Standard 2172-96
Heating Values
The
Dry BTU
,
Connected Dry BTU
,
Saturated BTU
and
Corrected Saturated BTU
are
calculated by equations 13-3 through 13-6.
Dry BTU
The energy content of the gas in BTU / cubic foot is calculated by equation 13-2.
Dry BTU / ft
3
=
Σ
Pn=1
[Conc
n *
BTU / ft
3n
] / 100
13-3
Where:
Dry BTU / ft
3
= Dry BTU content per cubic foot of sample gas
Conc
n
= Normalized concentration of component
BTU/ft
3n
= BTU value of component n
P
=
Number of components in the analysis
Corrected Dry BTU
Corrected Dry BTU / ft
3
= (Dry BTU / ft
3
) / Z
13-4
Saturated BTU
Sat BTU / ft
3
= Dry BTU / ft
3
x 0.9826
13-5
Where:
Sat BTU /ft
3
= Saturated BTU content per cubic foot of sample gas
Dry BTU /ft
3
= Dry BTU content per cubic foot of sample gas
Corrected Saturated BTU
Corrected Sat BTU / ft
3
= (Saturated BTU / ft
3
) / Z
13-6
Summary of Contents for AccuChrome MA2963
Page 21: ...Revision 4 8 12 15 2018 Figure 2 7 The Intrinsically Safe Barrier ...
Page 24: ...Revision 4 11 12 15 2018 Figure 2 10 4 Minute Analysis Flow Diagram ...
Page 58: ...Revision 4 45 12 15 2018 Figure 4 24 Chart Commands ...
Page 65: ...Revision 4 52 12 15 2018 A sample report is shown in Figure 4 32 Figure 4 32 Sample Report ...
Page 80: ...Revision 4 67 12 15 2018 Figure 5 10 Streams Setup ...
Page 104: ...Revision 4 91 12 15 2018 Figure 8 2 Enron Mode Short Integers ...
Page 140: ...Revision 4 127 12 15 2018 ...
Page 141: ...Revision 4 128 12 15 2018 15 2 Valve Maintenance Instructions ...
Page 142: ...Revision 4 129 12 15 2018 ...