Revision 4
68
12/15/2018
Run Name
- User defined
Run Type
- The stream type can be an Analysis, a Calibration or a Reference stream,
selected via the drop down menu. Selection of a Reference type selection presents the same
tab as a
Stream
type. The
Calibration
tab includes a button for
Manual Calibration
, which will
initiate a calibration sequence.
Action List Index
- used to indicate which
Action List
should be used (1 or 2).
Component Table Index
- used to indicate which
Component Table
should be used.
Allow Stream Switch
- used to indicate that the analyzer should switch streams during an
analysis. This allows the analyzer to purge the sample system with the upcoming stream for
a period of time prior to the analysis, so this is typically placed quite early in an analysis. The
stream that is switched to is defined either manually in the Analysis Control window, or
automatically in the Scheduling section of Sample Handling.
Purge Time
- used to indicate the amount of time the analyzer should purge the sample loop
prior to initializing a run definition
Reject Run from Archive
- If checked, the data from this stream will not be included in the
hourly or daily averages if any component high or low alarm is present for this stream.
The
Component Alarms
table is used to indicate the high and low value which should set off
an alarm. Enter the desired values and check the
Enable
box if the limits should be used
Additional tables for the Low and High Alarm Limits as well as Low and High Saturated Alarm
Limits can presented by clicking on the icons on the entries below the
Component Alarm
Limits
fields.
A typical table is presented as Figure 5-11 (the format of all of these tables is
identical).
Summary of Contents for AccuChrome MA2963
Page 21: ...Revision 4 8 12 15 2018 Figure 2 7 The Intrinsically Safe Barrier ...
Page 24: ...Revision 4 11 12 15 2018 Figure 2 10 4 Minute Analysis Flow Diagram ...
Page 58: ...Revision 4 45 12 15 2018 Figure 4 24 Chart Commands ...
Page 65: ...Revision 4 52 12 15 2018 A sample report is shown in Figure 4 32 Figure 4 32 Sample Report ...
Page 80: ...Revision 4 67 12 15 2018 Figure 5 10 Streams Setup ...
Page 104: ...Revision 4 91 12 15 2018 Figure 8 2 Enron Mode Short Integers ...
Page 140: ...Revision 4 127 12 15 2018 ...
Page 141: ...Revision 4 128 12 15 2018 15 2 Valve Maintenance Instructions ...
Page 142: ...Revision 4 129 12 15 2018 ...