Revision 5
122
August 2021
DO NOT touch the front face (light emitting side) of the lamp. Dirt, grease,
and human skin oils can affect the lamp spectral performance. Always
handle the lamp by the metal base.
DO NOT apply power to the lamp with the cover removed. Intense UV
radiation can cause severe eye damage.
11. Replace the two lamp retaining screws (5-40 x 3/8") and tighten using a 3/32" ball
driver passed though the holes on the top of the lamp enclosure.
12. Replace the two plastic plugs on the top of the lamp enclosure.
13. Reconnect the three wires from the lamp to the terminal strip. The wires are colour
coded on the terminal strip. Connect the lamp’s blue wire to the blue terminal, the red
wire to the red terminal, and the black wire to the black terminal. If the lamp has two
black wires instead of a blue and a black, connect one of the black wires to the blue
terminal and the other to the black terminal.
14. Reinstall the UV lamp enclosure cover. Tighten the cover cap screws.
15. Reconnect the AC terminal block fuse terminal on terminal block ACTS:28 by pressing
down and snapping it into place. Place the UV Lamp power switch to the ON position.
16. Close and latch the control cabinet door.
17. Ensure that the analyzer is still in backpurge mode.
18. Using the keypad, navigate to the
Spectrum
panel.
29. Optimize the integration period by using the F4 / F3 keys to highlight
Optimize Integ.
Time
, then press
↵
.
30. Observe the optimized integration time. If the
High Integration Period
warning has
cleared, no further work is required at this time. Proceed to the next step in this
procedure. If the
High Integration Period
warning is still present and the fibres have
already been tested as described in Section 7.3, however, it is likely that the cell
windows are dirty and/or damaged. Proceed to Section 7.5 – 7.6
31. Use the F2 / F1 keys to navigate to the
Analysis 1
panel. Perform a manual zero
calibration cycle by using the F4 / F3 keys to highlight
Manual Zero
, then press
↵
.
32. Place the analyzer into sampling mode by highlighting the
Sample
checkbox using the
F4 / F3 keys, then press
↵
. Alternatively, press the F6 key.
33. Once normal operation has been confirmed, return the
Online
/
Offline
control toggle
on the
Analysis 2
panel back to
Online
(green).
Summary of Contents for 943-TGS-CE
Page 137: ...Revision 5 137 August 2021 Figure 84 Oven Cabinet Details...
Page 138: ...Revision 5 138 August 2021 Figure 85 Control Cabinet Door Closed...
Page 139: ...Revision 5 139 August 2021 Figure 86 Control Cabinet Door Removed...
Page 140: ...Revision 5 140 August 2021 Figure 87 Power Steam Air Signals Connection Details...
Page 141: ...Revision 5 141 August 2021 Figure 88 AC Wiring Schematic...
Page 142: ...Revision 5 142 August 2021 Figure 89 DC Signals and Wiring Diagram...
Page 143: ...Revision 5 143 August 2021 Figure 90 Flow Diagram...
Page 144: ...Revision 5 144 August 2021 Figure 91 Control and Oven Cabinets Interconnect Wiring...