Revision 5
128
August 2021
13. Reconnect the AC Power to the oven heaters by pushing down and snapping closed
the oven heater fuse terminal cover (ACTS:25). The oven will begin heating to the set
point temperature.
14. Close and latch the oven enclosure cover.
15. Close and latch the control cabinet and oven cabinet doors.
16. Once the measurement cell temperature rises to >140°C, and the
Cell Temperature
fault has cleared, use the F2 / F1 keys to navigate to the
Analysis 1
panel. Perform a
manual zero calibration cycle by using the F4 / F3 keys to highlight
Manual Zero
, then
press
↵
..
17. Once the manual zero calibration cycle is complete, place the analyzer into sampling
mode by highlighting the
Sample
checkbox using the F4 / F3 keys, then press
↵
.
Alternatively, press the F6 key.
18. Once normal operation has been confirmed, return the
Online
/
Offline
control toggle
on the
Analysis 2
panel back to
Online
(green).
The oven can take up to two hours to reach operating temperature. While
the cell temperature is below the Low Cell Temperature fault set point, the
system will continue to back purge and maintain a fault indication to the
control room.
7.9
Steam Purging the Sample Probe
Steam Purging the Sample Probe is recommended
only when
a plug in the probe is suspected
and should only be performed as a last resort to avoid the possibility of steam or condensate
contamination in the sample cell, particularly on the cell windows. The residue of some boiler
water chemicals on the cell windows will cause a loss of spectrometer energy, potentially
resulting in
High Integration Period
warnings or even a fault condition in a worst-case scenario.
There are some indications potential warning signs that may indicate the potential need to
perform a steam purge of the probe. These include but are not limited to:
•
Absorbance spectrum and concentration readings remaining at zero even when the
analyzer is switched to sampling mode – this indicates that there may be a blockage
on the sample inlet side of the probe that is preventing sample from being drawn up
the probe and into the measurement cell.
•
Cell pressure increasing dramatically when analyzer is switched to sampling mode –
this indicates that there may be a blockage on the vent side of the probe that prevents
the aspirator drive air from entering the process pipe; this will cause a high sample cell
pressure.
•
Response time from zero to sample increases for a set aspirator drive air flow rate –
this indicates a partial blockage in the sample probe that is restricting the flow rate in
the probe and thus increasing the response time.
Prior to performing a steam purge, the method described in Section 3.4.4 using adjustments in
the zero air flow to clear obstructions in the probe and analyzer sampling system should be
attempted. Only if this method is unsuccessful should the steam purge procedure be performed.
Summary of Contents for 943-TGS-CE
Page 137: ...Revision 5 137 August 2021 Figure 84 Oven Cabinet Details...
Page 138: ...Revision 5 138 August 2021 Figure 85 Control Cabinet Door Closed...
Page 139: ...Revision 5 139 August 2021 Figure 86 Control Cabinet Door Removed...
Page 140: ...Revision 5 140 August 2021 Figure 87 Power Steam Air Signals Connection Details...
Page 141: ...Revision 5 141 August 2021 Figure 88 AC Wiring Schematic...
Page 142: ...Revision 5 142 August 2021 Figure 89 DC Signals and Wiring Diagram...
Page 143: ...Revision 5 143 August 2021 Figure 90 Flow Diagram...
Page 144: ...Revision 5 144 August 2021 Figure 91 Control and Oven Cabinets Interconnect Wiring...