Revision 5
25
August 2021
The oven cabinet lid is heavy and awkward. Removing the oven cabinet lid
is ideally a two person job, as it may be too awkward for a single person to
perform.
2. Insert the probe into the packing gland at the bottom of the oven indicated in Figure 4 until
it reaches the steam jacketed ball valve. It may be necessary to remove some tubing
connected to the sample cell in order to pass the sample probe through the packing gland.
Figure 4: Oven Cabinet Internal Layout
3. Open the steam jacketed ball valve and continue inserting the probe until the probe
flange contacts the packing gland.
Explosive, toxic and hot gases and molten sulfur may be released once the
stream jacketed ball valve is opened. All company and/or regulatory agency
safety procedures and rules should be adhered to. Breathing apparatus and
personal protective equipment (
ie.
gloves, coveralls and protective eye
wear) should be worn and a safety person(s) should be observing.
Do not
take unnecessary risks even if you believe the tailgas line to be at negative
pressure.
4. Tighten the packing gland and connect the tubing to the probe as shown in Figure 5.
Close the two block valves indicated (V2, V4) and open the other three block valves
(V1, V3, V5). Install the probe RTD in the location indicated.
Summary of Contents for 943-TGS-CE
Page 137: ...Revision 5 137 August 2021 Figure 84 Oven Cabinet Details...
Page 138: ...Revision 5 138 August 2021 Figure 85 Control Cabinet Door Closed...
Page 139: ...Revision 5 139 August 2021 Figure 86 Control Cabinet Door Removed...
Page 140: ...Revision 5 140 August 2021 Figure 87 Power Steam Air Signals Connection Details...
Page 141: ...Revision 5 141 August 2021 Figure 88 AC Wiring Schematic...
Page 142: ...Revision 5 142 August 2021 Figure 89 DC Signals and Wiring Diagram...
Page 143: ...Revision 5 143 August 2021 Figure 90 Flow Diagram...
Page 144: ...Revision 5 144 August 2021 Figure 91 Control and Oven Cabinets Interconnect Wiring...