Revision 5
56
August 2021
Below the Back Purge Indicator are two check boxes which can be used to manually set the
analyzer into back purge or sampling mode. To choose which mode is active, select the
desired mode using the F4 / F3 buttons and press
↵
to choose. Alternatively, pressing the
F6 key will switch to the currently inactive mode (i.e. pressing F6 while the analyzer is
sampling will cause it to switch to back purge mode and vice versa).
Prior to performing any maintenance, particularly if any sample system tubing is
to be disconnected, the analyzer should be placed into back purge mode and
purged for at least 5 minutes. This will prevent any toxic gases from remaining in
the sample system if tubing is to be disconnected.
If Sample is selected but the back purge indicator remains red, this indicates that a fault
condition likely exists. Once the fault condition(s) are cleared, the analyzer will automatically
return to sampling mode after first performing a zero calibration cycle. If the analyzer is set
to back purge mode while a fault condition exists, the analyzer will NOT automatically return
to sampling mode when the fault condition(s) are cleared.
4.4
Analysis 2 Panel
The
Analysis 2
panel shown in Figure 28 displays a variety of analysis information and system
operating parameters.
Figure 28: Analysis 2 Panel
4.4.1 Online / Offline Mode Toggle
At the top right of the
Analysis 2
panel is the
Online / Offline
Mode Toggle , as shown in
Figure 29.
Summary of Contents for 943-TGS-CE
Page 137: ...Revision 5 137 August 2021 Figure 84 Oven Cabinet Details...
Page 138: ...Revision 5 138 August 2021 Figure 85 Control Cabinet Door Closed...
Page 139: ...Revision 5 139 August 2021 Figure 86 Control Cabinet Door Removed...
Page 140: ...Revision 5 140 August 2021 Figure 87 Power Steam Air Signals Connection Details...
Page 141: ...Revision 5 141 August 2021 Figure 88 AC Wiring Schematic...
Page 142: ...Revision 5 142 August 2021 Figure 89 DC Signals and Wiring Diagram...
Page 143: ...Revision 5 143 August 2021 Figure 90 Flow Diagram...
Page 144: ...Revision 5 144 August 2021 Figure 91 Control and Oven Cabinets Interconnect Wiring...