Adjust
The offset adjustment for ECG and respiration waveforms can be performed on this menu.
The offset adjustment function can be used when baseline position of the waveform is not
properly displayed even though the waveform position has been adjusted.
Reference
For details of the waveform position adjustment, refer to the operation manual,
“6. Parameter Gain / Position Setup”.
CAUTION
Incorrect offset adjustment may cause failure of the equipment. As the
offset adjustment is performed at factory shipment, do not change the
setting unless it is necessary.
3)
6)
1)
4)
2)
5)
9)
12)
7)
10)
8)
11)
13)
14)
15)
1) 0 (ECG)
Sets the ECG offset value to “0”. This key will be highlighted with a red frame when the offset
value is set to “0”. (When the key is highlighted, the waveform will be also displayed with the
offset value of “0”.)
2) set (ECG)
Sets the offset value for the current ECG A/D value.
4) Offset Value = 3) A/D Value at the time the “2) set” key was pressed
´
(-1)
The offset value will be displayed at 4), and the “2) set” key will be highlighted with a red frame.
While the key is highlighted, the offset value is not still applied to the waveform. To finalize
the offset value setup, select “15) Flash Write” key. (Selecting the “2) set” key again will
cancel the setup, and the value will return to the previous offset value.)
3) ECG A/D Value
Displays the ECG A/D value without the offset adjustment.
4) Offset Value (ECG)
Displays the offset value set by selecting “1) 0” or “2) set” key.
5
-
11
Summary of Contents for DS-7001
Page 28: ...xxvi Blank Page ...
Page 40: ...1 6 Blank Page ...
Page 46: ...2 6 Blank Page ...
Page 57: ... SV 900 SV 300 Servo i Servo s CJ 501 Servo i Servo s CJ 502 SV 900 SV 300 CJ 500 3 11 ...
Page 78: ...4 16 Blank Page ...
Page 104: ...5 26 Blank Page ...
Page 126: ...7 12 Blank Page ...
Page 150: ...9 14 Blank Page ...