Equipment Status Alarm Message
The equipment status alarm message will be displayed when proper monitoring cannot be
performed. Some messages are displayed on the numeric data display area.
The message will be displayed according to the priority.
If more than one alarm are generated, the two higher priority alarms will be alternately
displayed.
CAUTION
While the “LEAD OFF” message is displayed, the HR alarm will not
generate. Leaving this condition unresolved may result in missing a
sudden change of the patient. Promptly check the electrodes when
this message is displayed.
Priority Message
Displayed
Color
Description
DEMO
DEMO
mode
High BATTERY
The remaining battery capacity is less than 10 minutes.
TELE. ER
Telemetry module error.
NO CARD
The CF card is not inserted.
Com Err
Communication error with the SpO
2
hardware
CVA
ECG waveform is superimposed on the respiration
waveform.
Sensor
Err
SpO
2
sensor failure.
PULSE?
Searching
pulse.
MOTION
Body motion is present during SpO
2
measurement.
NIBP ERR
NIBP measurement error.
(Cuff inflation pressure, measurement duration has
exceeded the upper limit.)
CF ERR
Detected error during CF card usage.
CF
FULL
Yellow
No space on the CF card to write the data.
CF REC
Recording waveform data on the CF card.
Low CF
INIT
Blue
Checking the inserted CF card.
7
-
10
Summary of Contents for DS-7001
Page 28: ...xxvi Blank Page ...
Page 40: ...1 6 Blank Page ...
Page 46: ...2 6 Blank Page ...
Page 57: ... SV 900 SV 300 Servo i Servo s CJ 501 Servo i Servo s CJ 502 SV 900 SV 300 CJ 500 3 11 ...
Page 78: ...4 16 Blank Page ...
Page 104: ...5 26 Blank Page ...
Page 126: ...7 12 Blank Page ...
Page 150: ...9 14 Blank Page ...