Periodic Check List (DS-7001 Series)
No.
.
Item
Inspection
Criteria
Judgment
Comment
1. Power Cable, Lead Cable, etc.
Perform visual check, cable
check for any abnormality.
o
OK /
o
NG
Cables,
Consumables,
Accessories
2. Operation Manual
Should be stored in
predetermined location.
o
OK /
o
NG
1. Scratch, crack, deformation, rust
No damage should be found.
o
OK /
o
NG
2. Peeled off label/panel, stain
Indication should be readable.
o
OK /
o
NG
Appeara
n
ce
External
Appearance
3. Key, lever, others
No damage should be found.
o
OK /
o
NG
Main Unit
CF Card
Should be smoothly inserted /
removed from the slot.
o
OK /
o
NG
Mec
h
an
ica
l
Control Key
Operation
All control keys should
properly operate.
o
OK /
o
NG
ECG
Waveform Size
10mm/mV ±10%
o
OK /
o
NG
Respiration Waveform
Size
10mm/
Ω
±10%
o
OK /
o
NG
SpO
2
Accuracy
95%
±2%
o
OK /
o
NG
NIBP
Operation
Should measure properly.
o
OK /
o
NG
Elec
tr
ic
al
Alarm Sound
Operation
Alarm sound should generate.
o
OK /
o
NG
LCD Unit
Check the date of previous
replacement.
Day Year Month
Should be within 10,000 hours
or 2 years.
o
OK /
o
NG
Lithium Battery
for Data
Backup
Check the date of previous
replacement.
Day Year Month
Should be within 6 years.
o
OK /
o
NG
P
eri
odi
c Repl
ac
em
ent
Pa
rt
Battery
Check the date of previous
replacement.
Day Year Month
Should be within 2 years.
Or, the number of
charge/discharge operation
should be within 300 times.
o
OK /
o
NG
1. Earth Leakage Current (Normal)
0.5mA or below
o
OK /
o
NG
2. Earth Leakage Current
(Single Failure)
1mA or below
o
OK /
o
NG
3. Exterior Leakage Current
(Normal)
0.1mA or below
o
OK /
o
NG
4. Exterior Leakage Current
(Single Failure)
0.5mA or below
o
OK /
o
NG
DC 0.01mA or below
o
OK /
o
NG
5. Patient Leakage Current
Ⅰ
(Normal)
AC 0.01mA or below
o
OK /
o
NG
DC 0.05mA or below
o
OK /
o
NG
6. Patient Leakage Current
Ⅰ
(Single Failure)
AC 0.05mA or below
o
OK /
o
NG
7. Patient Leakage Current
Ⅲ
(Single Failure)
0.05mA or below
o
OK /
o
NG
DC 0.01mA or below
o
OK /
o
NG
8. Patient Measurement Current
(Normal)
AC 0.01mA or below
o
OK /
o
NG
DC 0.05mA or below
o
OK /
o
NG
Leakage
Current
9. Patient Measurement Current
(Single Failure)
AC 0.05mA or below
o
OK /
o
NG
Between A-a1 (Primary Source and
Protective Earth)
Should withstand AC1000V for
one minute.
o
OK /
o
NG
Between B-d (Isolated Applied Part
and Signal I/O Part)
Should withstand AC1500V for
one minute.
o
OK /
o
NG
Safety
Withstand
Voltage
※
Between B-a (Isolated Applied Part
and Primary Source)
Should withstand AC3000V for
one minute.
o
OK /
o
NG
Overall Judgment
o
OK /
o
NG
Equipment
Bedside Monitor
Model Type
Serial Number
Installation
Place
Date of
Purchase
Date of
Inspection
Inspector
Approval
※
For the withstand voltage test between B-d and B-a, ECG input part should be used for isolated applied part.
4
-
15
Summary of Contents for DS-7001
Page 28: ...xxvi Blank Page ...
Page 40: ...1 6 Blank Page ...
Page 46: ...2 6 Blank Page ...
Page 57: ... SV 900 SV 300 Servo i Servo s CJ 501 Servo i Servo s CJ 502 SV 900 SV 300 CJ 500 3 11 ...
Page 78: ...4 16 Blank Page ...
Page 104: ...5 26 Blank Page ...
Page 126: ...7 12 Blank Page ...
Page 150: ...9 14 Blank Page ...