Periodic Replacement Parts
In this section, the components which need periodic replacement are listed.
About the Periodic Replacement Parts
To ensure reliability of safety, function, and performance of the DS-7001 system bedside
monitor, the following components must be replaced periodically.
Components Replacement
Period
LCD Unit
10,000 hours or more of continuous use,
or 2 years or more
Lithium Battery for Data Backup
(ER-3VC)
6 years
Battery Pack (10HR-AAC-4781)
2 years, or 300 times of charging / discharging
CAUTION
The periodic replacement parts must be replaced within specified
period.