Arterial Oxygen Saturation (SpO
2
)
Measurement :
2
Wavelength
Pulse
Wave
Method
Oxygen Saturation
Range :
1
~
100%
Resolution : 1%
Accuracy :
Adult
at
70
~
100%
±
2%
at
0
~
69%
not specified
Neonate
at
70
~
100%
±
3%
At
0
~
69%
not specified
Pulse Rate
Range :
20
~
250bpm
Accuracy :
±
3bpm
Sensor
Model Type : Nellcor
Ò
DURASENSOR, OXISENSOR III
Wavelength : 660nm (red, nominal value)
890nm (infrared, nominal value)
Telemetry
Transmitting Waveform
: ECG, RESP, Pulse Waveform
Transmitting Numeric Data
: HR, RR, SpO
2
, PR, NIBP
Status Data
: Lead Off, Probe Off, Low Battery, etc.
Medical Telemetry Type
: Type A
Occupied Frequency Range
: 8.5kHz and below
Communication :
Unidirectional
Oscillation
: Crystal controlled PLL synthesizer
Antenna Power
: 1.0mW
Radio Wave Type
: F7D
Transmission Frequency
: One wave within 420.0500
~
449.6625MHz
Leakage Power of Adjacent Channel :
-
40dB or below
Spurious Emission
: 2.5
m
W or below
Carrier Frequency Accuracy
:
±
4
´
10
‐
6
or below
8
-
4