Ventilator Connection
By connecting the ventilator to this monitor, ventilator information is input and ventilator
message such as communication status and alarm can be displayed on the screen for
monitoring.
This section explains the procedure to connect the DS-7001 to the ventilator and generate an
alarm.
When connecting the ventilator, check the corresponding version of the ventilator.
Ventilator
Corresponding Version
SV900 not
specified
SV300 not
specified
Servo-i v1.5
/
v2.0
Servo-s v2.0
PB-740 M
PB-760 H
PB-840 K
VELA
Main Software : MSP:VH2.07.00
IO Software
: IOP:V0.01.00
Evita 2 dura
04.14
Evita 4
04.14
Evita XL
05.10
E200 5.2.0
CAUTION
l
The ventilator operation should be performed by well-trained and
authorized personnel.
l
l
For connecting the DS-7001 and ventilator, use only the specified
connection cable.
Verify that the DS-7001 and the ventilator are properly connected.
l
When connecting the cable, verify that the main power of the
DS-7001 and the ventilator is OFF.
1 Connect the ventilator to the I/O port located on the backside of DS-7001 using
the optional connection cable.
Ventilator Connection
Cable
SV-900
CJ-500
SV-300 CJ-501
Servo-i / Servo-s
CJ-502
PB-740 / PB-760 / PB-840
CJ-504
VELA CJ-505
Evita 2 dura / Evita 4 / Evita XL
CJ-502
E200 CJ-506
3
-
10
Summary of Contents for DS-7001
Page 28: ...xxvi Blank Page ...
Page 40: ...1 6 Blank Page ...
Page 46: ...2 6 Blank Page ...
Page 57: ... SV 900 SV 300 Servo i Servo s CJ 501 Servo i Servo s CJ 502 SV 900 SV 300 CJ 500 3 11 ...
Page 78: ...4 16 Blank Page ...
Page 104: ...5 26 Blank Page ...
Page 126: ...7 12 Blank Page ...
Page 150: ...9 14 Blank Page ...