Evita 2 dura
】
Evita 2 dura
CJ-502
W ARNI NG
When connecting the Evita2dura / Evita4 / Evita XL ventilator, follow
the precautions below.
l
The serial port (RS-232C) setup of the ventilator should be as
follows. Refer to the service representative of the ventilator
manufacturer.
Protocol
: Medibus
Baud Rate : 19200bit/s
Data Bit
: 8bit
Parity Bit
: Even
Stop Bit
: 1bit
l
The DS-7001 will not correspond to the following alarm generated at
the ventilator.
O
2
monitoring disabled alarm
CO
2
alarm disabled alarm
Oximeter alarm disabled alarm
Neo. volume measurement inoperable alarm
Minute volume alarm disabled alarm
Minute volume alarm low off alarm
Tidal volume alarm high off alarm
Apnea alarm off alarm
Nebulizer active alarm
3
-
14