Precautions about the DS-7001
DANGER
When connecting to other device, contact Fukuda Denshi service
representative.
Danger such as electric shock may result to the patient and
operator.
WARNING
l
l
l
l
Do not connect unit or cable not authorized by Fukuda Denshi to
any I/O connector. If done so by mistake, the device can not
deliver its maximum performance and the connected units may be
damaged, resulting in a safety hazard.
l
Use only the accompanying 3-way AC power cable. Use of other
cables may result in electric shock to the patient and the operator.
l
The power cable must be plugged into 230V AC hospital grade
outlet.
Use only the accompanying 3-way AC power cable to assure
ground reliability. Use a power outlet that can supply sufficient
power to the device.
l
When using multiple ME devices simultaneously, equipotential
grounding must be conducted to avoid electrical potential difference
between the devices. Presence of any electrical potential
difference may result in electric shock to the patient and the
operator.
The battery installation will be performed by our service
representative. Users should not attempt the procedure as it may
result in electric shock or malfunction to the equipment.
If the upper/lower alarm threshold is set to OFF, alarm will not
generate even if all alarm is set to ON. Pay attention when setting
the threshold OFF.
l
Even if the individual parameter alarm is set to ON, all the alarm will
not generate during the alarm suspend time.
Check the patient’s condition frequently.
l
When using electrosurgical instrument with this equipment, verify
ground plate attachment, electrode position, ground cable
connection, etc.
A high frequency energy generated by the electrosurgical
instrument may cause burn injury at the electrode site, damage the
equipment, or cause electric shock.
l
Do not touch the DS-7001 during defibrillation.
Electric shock may result to the patient or the operator.
l
Do not disassemble/remodel the equipment.
Electric shock may result.
l
Set the correct patient type on the admit menu.
This selection will affect the measurement accuracy of QRS
detection, cuff inflation value of NIBP measurement, and respiration
filter.
l
Set the pacemaker use (Yes/No) correctly on the admit menu.
This will affect accuracy of QRS detection. The pacemaker pulse
may be misdetected as QRS wave causing to overlook pacing
failure.
When measuring the SpO
2
of patient with high fever or peripheral
circulatory insufficiency, check the sensor attachment periodically
and change the attachment site. The temperature of attachment
site will rise 2
~
3
°
C due to the sensor heat which may result in burn
injury.
xi
Summary of Contents for DS-7001
Page 28: ...xxvi Blank Page ...
Page 40: ...1 6 Blank Page ...
Page 46: ...2 6 Blank Page ...
Page 57: ... SV 900 SV 300 Servo i Servo s CJ 501 Servo i Servo s CJ 502 SV 900 SV 300 CJ 500 3 11 ...
Page 78: ...4 16 Blank Page ...
Page 104: ...5 26 Blank Page ...
Page 126: ...7 12 Blank Page ...
Page 150: ...9 14 Blank Page ...