W ARNI NG
l
Use only the accompanying 3-way AC power cable. Use of other
cables may result in electric shock to the patient and the operator.
l
The power cable must be plugged into 230V AC hospital grade
outlet.
l
When using multiple ME devices simultaneously, equipotential
grounding must be conducted to avoid electrical potential
difference between the devices. Presence of any electrical
potential difference may result in electric shock to the patient and
the operator.
NOTE
Equipotential grounding
When connecting multiple devices, electrical potential difference may be
generated between the devices. This may result in electric shock to the
patient who is connected to this device. Pay special attention for use in
the operating room, ICU, CCU, Cardiac Catheter Laboratory, and
Cardiovascular X-ray room. To avoid such electrical potential
difference, use the accessory ground cable to connect each device’s
equipotential terminal to the same ground terminal. This is called
equipotential grounding.
For Use with the Battery
Using the battery for operating the DS-7001 allows monitoring at places where AC power source
is not available such as during in-hospital patient transportation.
One battery is used for the DS-7001. When fully charged, the DS-7001 is capable to operate
for about 60 minutes.
Reference
For procedure to replace the battery, refer to “6. Periodic Replacement”.
Also, the DS-7001 will automatically switch to battery operation for the following cases.
·
When a sudden electricity failure occurs.
·
When the AC power cable is accidentally disconnected.
CAUTION
l
The battery operating duration will shorten depending on the
number of NIBP measurement performed, etc.
l
When using a new battery, or a battery which was stored for a long
period of time, the operating duration may be short at beginning of
usage. In such case, repeating charging and battery operation for
2 or 3 times will bring out the original performance.
W ARNI NG
The battery installation will be performed by our service representative.
Users should not attempt the procedure as it may result in electric
shock or malfunction to the equipment.
3
-
4
Summary of Contents for DS-7001
Page 28: ...xxvi Blank Page ...
Page 40: ...1 6 Blank Page ...
Page 46: ...2 6 Blank Page ...
Page 57: ... SV 900 SV 300 Servo i Servo s CJ 501 Servo i Servo s CJ 502 SV 900 SV 300 CJ 500 3 11 ...
Page 78: ...4 16 Blank Page ...
Page 104: ...5 26 Blank Page ...
Page 126: ...7 12 Blank Page ...
Page 150: ...9 14 Blank Page ...