“Com Err” is displayed in the SpO
2
numeric data area, and the measurement is
ceased.
Cause : The hardware is not properly operating due to static electricity, etc.
Solution : Wait for about 2 minutes until the “Com Err” message disappears.
If the message does not disappear, restart the system.
If the “Com Err” message is displayed again, cease the measurement and check the
SpO
2
board status on the test menu.
Reference
For details of the test menu, refer to “5. Function Test”.
“Sensor Err” is displayed in the SpO
2
numeric data area, and the measurement
is ceased.
Cause :
Sensor
is
defective.
Solution : Replace the sensor.
Alarm
Alarm sound does not generate. Alarm is not displayed.
Cause 1 : The alarm is suspended.
Solution : While the alarm is suspended, the message “ALARM SUSPEND” is displayed and
alarm will not generate even if the individual parameter alarm is set to ON. When
the suspend time completes, the system alarm automatically resets to ON. To set
the system alarm ON, cancel the alarm suspend function on the alarm setup menu.
Cause 2 : The individual parameter alarm is set to OFF.
Solution : Even if “All Alarm” is set to ON, if individual parameter alarm is set to OFF, the
alarm will not generate. Turn ON the alarm on the alarm setup menu for each
parameter.
Reference
For the parameter which the alarm is set OFF, the alarm off symbol “
” is displayed in the
numeric display area
Cause 3 : Individual parameter alarm limits are not set.
Solution : Set the individual parameter alarm limits.
Reference
Even if the individual parameter alarm is
ON , if the parameter limits are
OFF , the alarm
off symbol “
” is displayed in the numeric display area.
7
-
5
Summary of Contents for DS-7001
Page 28: ...xxvi Blank Page ...
Page 40: ...1 6 Blank Page ...
Page 46: ...2 6 Blank Page ...
Page 57: ... SV 900 SV 300 Servo i Servo s CJ 501 Servo i Servo s CJ 502 SV 900 SV 300 CJ 500 3 11 ...
Page 78: ...4 16 Blank Page ...
Page 104: ...5 26 Blank Page ...
Page 126: ...7 12 Blank Page ...
Page 150: ...9 14 Blank Page ...