Analog 2
This screen displays the status of the data backup lithium battery and other batteries.
The A/D converted data will be displayed.
Batt L
Displays the voltage of the data backup lithium battery.
If the voltage is 2.0000V or below, the lithium battery will be judged as depleted, and ”SRAM
Battery NG” will be displayed at startup.
The actual voltage can be calculated from the following equation.
Batt L value
´
1.5 = Actual Voltage
Batt D
Displays the battery voltage.
The actual voltage can be calculated from the following equation.
Batt D value
´
5.55 = Actual Voltage
sum
Displays the checksum status.
DIP
Displays the DIP switch status. “1” is ON, and “0” is OFF.
LD
Displays telemetry unit status.
TTL
Displays TTL input status.
HLX
Reserved function. Not used.
5
-
8