Precautions for Ventilator Monitoring
WARNING
l
The DS-7001 may not be able to properly receive the alarm
information from the ventilator due to causes such as communication
error, using environment, usage procedure, etc. Therefore, always
check the devices (ventilator, communication status of ventilator and
DS-7001, etc.) and safety of the patient. If any abnormality is found
on the devices or patient, take appropriate actions such as ceasing
the operation of the devices in the safest way for the patient.
l
The ventilator operation should be performed by well-trained and
authorized personnel.
l
If the DS-7001 does not generate an alarm even though the
ventilator is generating an alarm, or if any other malfunction occurs,
immediately check the ventilator, monitor, and the cable, and replace
the cable if necessary.
If malfunction persists, cease using the monitor.
l
The alarm generation on the DS-7001 is not assured if the alarm
other than specified generates at the ventilator
(
SV-900, SV-300,
Servo-i/s, PB-740/760/860, VELA
)
.
See For details of “specified alarm”, refer to
WARNING on the
operation manual, P3-29 “Ventilator Monitoring”.
l
When connecting the ventilator to the monitor, using them under
adverse environmental condition will not only disallow the devices to
deliver their maximum performance, but also the devices may be
damaged and safety cannot be ensured.
l
When transmitting the ventilator information to the central monitor
using the medical telemetry, read the “Precautions for Safe
Operation of Medical Telemetry” in the Preface.
l
l
l
When monitoring the ventilator information on the central monitor, use
the DS-7600 system or DS-5700 Central Monitor. Do not use the
DS-5800N/NX/NX
MB
Central Monitor. For use of other central
monitors, contact our service representative.
l
When monitoring ventilator information, pulse oximetry (SpO
2
) can
be also simultaneously monitored.
l
For connecting the DS-7001 and ventilator, use only the specified
connection cable.
Precautions about Evita2dura / Evita4 / EvitaXL
·
The Evita2dura / Evita4 / EvitaXL acquires alarm information from the
serial port. The ventilator alarm that cannot be acquired from the serial
port is not guaranteed.
For corresponding alarm, refer to the service representative of the
ventilator manufacturer.
The DS-7001 will not correspond to the following alarm generated at
Evita2dura / Evita4 / EvitaXL.
O
2
monitoring disabled alarm
CO
2
alarm disabled alarm
Oximeter alarm disabled alarm
Neo. volume measurement inoperable alarm
Minute volume alarm disabled alarm
Minute volume alarm low off alarm
Tidal volume alarm high off alarm
Apnea alarm off alarm
Nebulizer active alarm
xvi
Summary of Contents for DS-7001
Page 28: ...xxvi Blank Page ...
Page 40: ...1 6 Blank Page ...
Page 46: ...2 6 Blank Page ...
Page 57: ... SV 900 SV 300 Servo i Servo s CJ 501 Servo i Servo s CJ 502 SV 900 SV 300 CJ 500 3 11 ...
Page 78: ...4 16 Blank Page ...
Page 104: ...5 26 Blank Page ...
Page 126: ...7 12 Blank Page ...
Page 150: ...9 14 Blank Page ...