Ambulatory Blood Pressure 

Report Generator

90239A

Operations Manual

070-0399-03 Rev. F

Summary of Contents for Spacelabs Medical 90239A

Page 1: ...more time to care Ambulatory Blood Pressure Report Generator 90239A Operations Manual 070 0399 03 Rev F ...

Page 2: ...lified technical personnel to repair those parts of the equipment which are classified by Spacelabs Medical as field repairable Spacelabs Medical is committed to providing comprehensive customer support beginning with your initial inquiry through purchase training and service for the life of your Spacelabs Medical equipment CORPORATE OFFICES U S A Spacelabs Medical Inc 5150 220th Ave SE Issaquah W...

Page 3: ... External Printer Configuration 1 5 Configuration Menu Setup 1 6 Report Generator Operation 1 9 Copying ABP Monitor Data to a Data Card 1 9 Printing Patient Data 1 10 Printing and Copying a Patient Report 1 11 Correcting Printer Paper Problems 1 12 Routine Maintenance 1 12 Performance Verification 1 12 ABP Monitor 1 13 Sample Printouts 1 13 Internal Printer Samples 1 14 External Printer Samples 1 ...

Page 4: ......

Page 5: ...can quickly access just those parameters that you want to alter The menu items are self explanatory requiring you simply to enter a numeric value The report generator software can be upgraded with a ROM card that fits into a card slot below the control panel Contact your sales representative for details Data Card Option The Data Card option P N 040 0549 00 permits multiple blood pressure reports t...

Page 6: ...Figure 1 5 Figure 1 1 90239A ABP Report Generator ABP monitor placed in chute of 90239A Report Generator for data transfer Label providing definitions for menu items User control panel Card slot for Data Card option Removable plastic cover to access printer paper Figure 1 2 90239A rear view Input connector for power adapter 9 VDC 800 mA 110 V P N 655 0636 00 230 V P N 119 0125 00 Serial port for c...

Page 7: ...ght are as follows PRINT HELP LINEFEED INITIALIZE AND PRINT REPORT PRINT HELP Press this key once to print brief definitions of the four keys on the right When printing is complete press it again to print brief definitions of the five keys on the left This key also prints specified menu items refer to Changing Menu Configurations on page 1 8 Press this key during printing to abort printing the key...

Page 8: ...other brands of paper are used 1 Ensure that the print head is positioned at the far left If not apply electrical power to the report generator The print head will now make one or more passes across the print area When it holds its position against the left side for several seconds disconnect electrical power 2 Remove the plastic cover over the cradle for the thermal paper and orient the roll so t...

Page 9: ...ctor and printer cable Refer to Figure 1 5 below Figure 1 5 External printer configuration Either of these two printers can be used Okidata Microline 320 or 100 compatible IBM ProPrinter or 100 compatible Two additional menu items now appear on the 90239A to support use of an external printer Menu Item Selections Available Description 37 00 Messages printed on internal printer 01 Messages printed ...

Page 10: ...hours 23 00 04 Specify time interval between each measurement during waking hours 20 minutes 05 Is a tone to sound 5 seconds before each programmed reading taken during waking hours yes or no Yes 06 Specify the time interval between each measurement during sleeping hours 60 minutes 07 Is a reading to be displayed on the monitor after each measurement yes or no Yes 08 Is cuff pressure to be display...

Page 11: ... 40 30 Specify the auto edit limit for maximum pulse pressure displayed on monitor as value divided by 10 150 31 Specify the auto edit limit for minimum pulse pressure displayed on the monitor as value divided by 10 20 32 Specify the auto edit limit for maximum heart rate displayed on monitor as value divided by 10 200 33 Specify the auto edit limit for minimum heart rate displayed on monitor as v...

Page 12: ...see that the numbers scroll to 23 denoting 11 00 P M and then start over at 00 The menu is always in a 24 hour format 4 Press ENTER to change the waking hour to the displayed value The new value will stop flashing when it has been saved to memory Note The ABP monitor will continue to use the previous wake up setting of 6 00 A M if the ENTER key is not pressed Printing the Configuration Menu A prin...

Page 13: ...lization the unprinted data will be erased OR Press ENTER to abort the initialization process In this case the report generator prints Monitor initialization aborted Copying ABP Monitor Data to a Data Card To copy patient data from an ABP monitor to a data card optional 1 Insert a data card into the card slot on the front of the ABP Report Generator refer to Figure 1 1 item Ensure that the data ke...

Page 14: ...ata table one table for each 24 hours 7 Edited data table one table for each 24 hours 8 Raw data pressure graph one graph for each 24 hours 9 Raw data heart rate graph one graph for each 24 hours Data that was obtained when the patient manually started a reading via the ABP monitor s START STOP switch will be tagged with an m for manual reading Data that was obtained on a second try after the ABP ...

Page 15: ...t of the patient data reports to be printed Additional patient data reports can be added to the print queue by selecting their numbers and pressing ENTER 4 To verify the report numbers in the print queue prior to printing use the left UP or DOWN arrow key to select menu item 40 then use the right UP or DOWN arrow key to select 00 Press ENTER to print a list of all report numbers that are in the qu...

Page 16: ...l the remaining paper out only in the direction indicated in Figure 1 6 Figure 1 6 Clearing paper jams Routine Maintenance The following maintenance activities should be done on an as needed basis Wipe all external surfaces with a damp cloth Ensure that no dust lint etc obstructs the infrared window in the bottom of the chute Remove the printer cover and vacuum out accumulated dirt and paper fiber...

Page 17: ... unsuccessful self test will display EC25 indicating that the ABP monitor s software program has been corrupted Place the ABP monitor in the chute and the report generator will reload the software program When the reloading process is complete the ABP monitor will display 0000 and the report generator prints the message Monitor software successfully installed in monitor Note Event Code EC numbers ...

Page 18: ...Ambulatory Blood Pressure Report Generator 1 14 Internal Printer Samples ...

Page 19: ...Ambulatory Blood Pressure Report Generator 1 15 Report Generator Internal Printer Samples continued ...

Page 20: ...Ambulatory Blood Pressure Report Generator 1 16 External Printer Samples ...

Page 21: ...Ambulatory Blood Pressure Report Generator 1 17 Report Generator External Printer Samples continued New Graphics to be added ...

Page 22: ...Ambulatory Blood Pressure Report Generator 1 18 External Printer Samples continued New Graphics to be added ...

Page 23: ...Ambulatory Blood Pressure Report Generator 1 19 Report Generator External Printer Samples continued ...

Page 24: ...Ambulatory Blood Pressure Report Generator 1 20 External Printer Samples continued ...

Page 25: ...NCTIONS Key Mouse connection RECORD Key START STOP Key NORMAL SCREEN Key START STOP MONITOR SETUP Key STOP or CANCEL Key ALARMS Key CONTINUE Key PREVIOUS MENU Key ENTER Key ON Power Connection to Mains OFF Power Disconnection from Mains ON Position for Push Button Power Switch OFF Position for Push Button Power Switch On Direction ON OFF Television Video Display Video Output ON Part of the Instrum...

Page 26: ...Indicator Remote Control Indicator Local Control PRINT REPORT Key Indicator Out of Paper Partial ON OFF Recorder Paper Normal Screen Return to Prior Menu Clock Time Setting Key TREND TIMER Key HELP Explain Prior Screen Key Keypad Activate Recorder for Graphics Indoor Use Only START NIBP Key Auto Mode NIBP Output No Output Terminated Symbol Description Symbol Description 1 2 3 1 2 3 ...

Page 27: ... Port 2 External marker push button connection SDLC Port Arterial Pulse Electrocardiograph or Defibrillator Synchronization Gas Exhaust Foot Switch Enlarge Zoom Delete PCMCIA Card Event Keep Dry Fragile handle with care Environmental Shipping Storage Altitude Limitations This Way Up Environmental Shipping Storage Temperature Limitations Environmental Shipping Storage Humidity Limitations Symbol De...

Page 28: ...appropriate battery Replace only with the appropriate battery signs may be reversed All batteries should be disposed of properly to protect the environment Lithium batteries should be fully discharged before disposal Batteries such as lead acid Pb and nickel cadmium Ni Cd must be recycled Please follow your internal procedures and or local provincial laws regarding disposal or recycling Caution ha...

Page 29: ...solated floating patient applied part which contains an adequate degree of protection against electric shock and is defibrillator proof IEC 601 1 Type BF equipment The unit displaying this symbol contains an F type isolated floating patient applied part providing an adequate degree of protection against electric shock IEC 601 1 Type CF equipment The unit displaying this symbol contains an F type i...

Page 30: ...cription WARNING Warning About Potential Danger to Human Beings CAUTION Caution About Potential Danger to a Device Noninvasive Blood Pressure NIBP Neonate Fetal Monitor Connection Analog Fetal Monitor Connection RS232 Digital Physiological Monitor Connection RS232 Digital Happy Face Sad Face Magnifying Glass Compression File Cabinet List of Rooms Arrows Printer Recycle Service Message Radio transm...

Page 31: ...t DIA dia Diastolic ECG ecg Electrocardiogram EEG eeg Electroencephalogram EMG emg Electromyogram ESIS Electrosurgical Interference Suppression EXT External FECG Fetal Electrocardiogram FHR1 FHR2 Fetal Heart Rate Channel 1 Fetal Heart Rate Channel 2 GND gnd Patient Isolated Ground HLO hlo High Level Output Multiview Multi Lead Electrocardiogram NIBP nibp Noninvasive Blood Pressure N2O Nitrous Oxid...

Page 32: ...Report Generator 2 8 SYS sys Systolic T1 T2 T3 T4 Temperature 1 Temperature 2 Temperature 3 Temperature 4 TEMP temp Temperature UA Uterine Activity or Umbilical Artery VAC Vacuum Connection Symbol Description Symbol Description ...