Ambulatory Blood Pressure Report Generator
1-13
Report Generator
ABP Monitor
The ABP monitor performs a self-test each time its power switch is turned ON. Following a
successful self-test, the monitor displays its software version number (0209 or a similar value) and
then displays the current time. If the ABP monitor is inserted into the chute of the report generator
at this time, it flashes a series of numbers and then displays “0000.”
Note:
If the ABP monitor maintains a display of “9999,” the report generator is not powered ON.
An unsuccessful self-test will display “EC25,” indicating that the ABP monitor’s software program
has been corrupted. Place the ABP monitor in the chute, and the report generator will reload the
software program. When the reloading process is complete, the ABP monitor will display “0000”
and the report generator prints the message “Monitor software successfully installed in monitor.”
Note:
• Event Code (EC) numbers are defined in the 90207, 90217 Ambulatory Blood Pressure
Monitors Operations Manual (P/N 070-0137-xx).
Sample Printouts
The following pages show examples of internal and external printouts.
Note:
For explanations of Event Code (EC) numbers, refer to the 90207, 90217 Ambulatory Blood
Pressure Monitors Operations Manual (P/N 070-0137-xx).