SS1807-N002
Failure Analysi
-
36/64
Display
error
code
Classification
level
Display message
(occurrence, response)
Display message
(guidance to
error)
Estimated cause
Trigger of message display
Service response
9020
Warning
Some image files are not
yet transferred.
Transfer these files to
Storage server again?
Under the setting for non-use of
media storage at transfer during
capture, communication failure to
the storage server occurred, and
transfer of all the image files was
not completed. Under this
condition, examination
termination was executed.
Same as on the left
Dialog selection→
(1) Check connection to the
storage server, and select
“Transfer.”
(2) If it is safe to delete the image
files, select “Delete.”
In case of repetitive failures,
check the network cable and the
network device
9021
Warning
Could not send
termination notification of
MPPS.
Send it again?
Communication failure to MPPS
server
Same as on the left
Check network connection and
setting.
9022
Warning
Obtaining worklist failed.
Cause: MWM
server is not
connected to the
network or
network setting is
incorrect.
The server is not connected to
the network or something wrong
with network setting
Same as on the left
Check network connection and
setting.
9023
Warning
Communication with
Storage server is not
established.
Create image file?
Cause: Storage
server is not
connected to the
network, or
network setting is
incorrect.
The server is not connected to
the network or something wrong
with network setting
When an event of examination
start preparation was received
Check network connection and
setting.
9024
Information
Transferring image files to
Storage server.
Please wait for a while.
Image transfer when examination
is finished.
When storage transfer was
performed when examination
was finished.
Nothing special
9025
Information
Creating DICOMDIR file
succeeded.
When writing of the DICOMDIR
index file to the external memory
is successful.
Same as on the left
9026
Information
Deleting DICOM files.
Please wait for a while.
Automatic deletion when
examination is finished.
When auto deletion is performed
at the time of examination end
Nothing special
9027
Information
Number of shots during
examination is limited.
Note that shooting may
be disabled during
examination.
To release the
limit, enable
external memory
backup function
and insert external
memory.
Inspection was started while not
saving to the external memory
during batch transfer.
Same as on the left
Review DICOM setting for users.
9028
Information
Exporting DICOMDIR is
not executed because no
DICOM images exist.
Cause: No DICOM
files exist in
Internal Memory.
A DICOM file was exported while
no DICOM file was saved on the
external memory.
Same as on the left
Nothing special
Summary of Contents for EP-6000
Page 1: ...Processor EP 6000 FV693A Service Manual SR1807 N002 Ver 1 Oct 2018 ...
Page 5: ...SS1807 N002 General Table of Contents 1 1 General Table of Contents ...
Page 13: ...SS1807 N002 Caution in Safety 1 12 Caution in Safety ...
Page 25: ...SS1807 N002 Product Specifications 1 11 Product Specifications ...
Page 36: ...SS1807 N002 Instruction of System 1 106 Instruction of System ...
Page 133: ...SS1807 N002 Instruction of System 98 106 ELC PCB Patient PCB APC PCB APC PCB APC PCB DC Pump ...
Page 139: ...SS1807 N002 Instruction of System 104 106 9 4 Outline of PCB roles ...
Page 142: ...SS1807 N002 Failure Analysis 1 64 Failure Analysis ...
Page 206: ...SS1807 N002 Checkup Replacement and Adjustment 1 137 Checkup Replacement and Adjustment ...
Page 343: ...SS1807 N002 Service Parts List 1 19 Service Parts List ...
Page 352: ...SS1807 N002 Service Parts List 10 19 2 2 2 4 1 3 3 2 3 3 6 5 ...
Page 356: ...SS1807 N002 Service Parts List 14 19 1 3 3 4 1 2 1 3 X 4 1 3 3 4 3 3 ...
Page 358: ...SS1807 N002 Service Parts List 16 19 3 1 2 1 ...
Page 360: ...SS1807 N002 Service Parts List 18 19 Fig 09 5 4 2 3 1 ...
Page 362: ...SS1807 N002 Periodical Maintenance 1 15 Periodical Maintenance ...
Page 377: ...SS1807 N002 Installation 1 68 Installation ...
Page 445: ...SS1807 N002 ...