SS1807-N002
Instruction of System
-
39/106
Menu item
Setting
Default
Description
Remarks
R-gain
0 to 960
The current values of the connected
Endoscope are displayed.
Placing the cursor on the item and pressing
the Enter key display the sider. Specify a
value with the ← and → keys and set using
the Enter key
G-gain
0 to 960
B-gain
0 to 960
Cancel Adjustment
Values after manual adjustment are returned
to those before automatic adjustment.
Reset to Factory Default Setting
R-gain, G-gain, B-gain, values are reset to
the factory defaults.
Execute Writing
When this item is selected and the Enter key
is pressed, the scope color adjustment
values are overwritten onto the EEROM
[Note] When the adjustment values are
written onto the EEROM, these values
are used for subsequent operations.
[Note] When “Save and Exit” is selected
without selecting "Execute Writing",
the color adjustment value is only
applied to the current examination.
Save and
Exit
When this item is selected and the Enter key is pressed, the current
settings are finalized and operation returns to the System Configuration
screen.
[Note] Pressing the Esc key without selecting “Save and Exit” canceled
the current setting and operation returns to the System
Configuration screen.
<Scope Info>
Information on the connected Endoscope is display (Display only).
Displayed items vary depending on the version of the scope CPU on the connected
Endoscope and scope series.
A: 530 series scope (scope CPU of Ver. 2.0 or later). 550 series scope
B: 530 series scope (scope CPU earlier than Ver. 2.0).
Information
Value
Description
A
B
Type
(16
characters)
Type of connected Endoscope
Displayed
Not
displayed
Serial
(16
characters)
Serial number of the connected
Endoscope
Displayed
[Note]
Not
displayed
ID
(3
characters)
Scope ID of the connected Endoscope
Displayed
Displayed
Usage
Count
(10
characters)
Number of operation times of the
connected Endoscope
Displayed
Not
displayed
Repair
History
(6
characters)
Repair history code
Displayed
Not
displayed
Repair Date
(4,2 and 4
characters)
Date of repair (year/month/day)
Displayed
Not
displayed
Factory
Shipment
Date
(4,2 and 4
characters)
Factory Shipment Date (year/month/day)
of the Endoscope
Displayed
[Note]
Not
displayed
[Note] If the scope data supporting the EP-6000 is not written in the connected Endoscope, the
scope data possessed by the processor is displayed in place of the scope data
possessed by the Endoscope.
Accordingly, the displayed data does not match with the scope data of the connected
Endoscope.
Summary of Contents for EP-6000
Page 1: ...Processor EP 6000 FV693A Service Manual SR1807 N002 Ver 1 Oct 2018 ...
Page 5: ...SS1807 N002 General Table of Contents 1 1 General Table of Contents ...
Page 13: ...SS1807 N002 Caution in Safety 1 12 Caution in Safety ...
Page 25: ...SS1807 N002 Product Specifications 1 11 Product Specifications ...
Page 36: ...SS1807 N002 Instruction of System 1 106 Instruction of System ...
Page 133: ...SS1807 N002 Instruction of System 98 106 ELC PCB Patient PCB APC PCB APC PCB APC PCB DC Pump ...
Page 139: ...SS1807 N002 Instruction of System 104 106 9 4 Outline of PCB roles ...
Page 142: ...SS1807 N002 Failure Analysis 1 64 Failure Analysis ...
Page 206: ...SS1807 N002 Checkup Replacement and Adjustment 1 137 Checkup Replacement and Adjustment ...
Page 343: ...SS1807 N002 Service Parts List 1 19 Service Parts List ...
Page 352: ...SS1807 N002 Service Parts List 10 19 2 2 2 4 1 3 3 2 3 3 6 5 ...
Page 356: ...SS1807 N002 Service Parts List 14 19 1 3 3 4 1 2 1 3 X 4 1 3 3 4 3 3 ...
Page 358: ...SS1807 N002 Service Parts List 16 19 3 1 2 1 ...
Page 360: ...SS1807 N002 Service Parts List 18 19 Fig 09 5 4 2 3 1 ...
Page 362: ...SS1807 N002 Periodical Maintenance 1 15 Periodical Maintenance ...
Page 377: ...SS1807 N002 Installation 1 68 Installation ...
Page 445: ...SS1807 N002 ...