Menu 4: Physical units (UNIT)
The FCX-AII pressure transmitters
intended for use in nuclear power
plants are delivered configured
with an enabled write protection.
The configuration of the physical units on the
device must be done before the write protection
is lifted according to the provision specified on
page 35. After all the operations are complete,
the write protection must be replaced.
26 physical units can be selected by the user
(see below).
Note
: The units marked < > are configurable
only for the absolute pressure transmitters.
• To change the unit, press <CHNG> in
screen 1 . Screen 2 appears, allowing the
transmitter unit to be modified.
• Select the unit using the <INC> or <DEC>
keys on screen 2 .
• Screen 3 appears for confirmation of the
change.
• Screen 4 for the registration of the unit and
then screen 1 returns.
At each stage, it is possible to go back through
the key <CL>
• To display the menu <RANGE LIMIT>, press
the <INC> key as indicated on screen 1 .
If screen 5 appears (otherwise
reset screen 5 ) during the
modification of the unit. The latter
may not be taken into account
due to the scale selected.
In this case, press <CL> and change the unit.
< UNIT CHANG >
4-1: UNIT CHANGE
kPa
< INC > < ENT > < CL >
4-1: UNIT CHANGE
CHNG OK? < ENT/CL >
4-1: UNIT WRITE
< INC >
4: UNIT
5 : RANGE LIMIT
kPa
< CHNG >
(NEXT MPa)
(NEXT MPa)
kPa
kPa
4-1: UNIT WRITE
UNIT < CL >
NOT SUITABLE
Not suitable unit display
Pa
3 :
SERIAL N
CL
CL
CL
INC
INC
CHNG
ENT
ENT
1
2
3
4
INC
5
DEC
INC
mmH O
2
cmH O
2
mH O
2
g/cm
2
Pa
inH O
2
psi
ftH O
2
< Torr >
kg/cm
2
< atm >
hPa
kPa
MPa
mbar
bar
mmAq
mmHg
cmAq
mAq
mmWC
cmWC
mWC
inHg
cmHg
mHg
Important
Important