3
TN4FCXA2
j
-EDF-
E
CAUTION:
Rotation of the whole upper part:
The whole upper part (h electronics) can be rotated by 90° steps to the left or to the right
by removing the 2 socket head cap screws (M6 x 12)
If the entire assembly must be turned more than 90°, or if it
is not known if the position has already been amended since
the delivery by Fuji, you must remove the electronics of the
unit and disconnect the flexible ribbon cable connecting the
electronics at the measuring cell to rotate the housing.
If necessary, modify the position of the flexible ribbon cable
connecting the electronics and the measuring cell, then refit
the different parts.
The non-observance of these instructions may lead to the
deterioration of the flexible ribbon cable, which would not be
covered by the manufacturer’s warranty.
To find out which models are qualified K3A and K3-ad, refer to the latest versions of
the revision of the technical specifications of each model.
Example, for the DIFFERENTIAL AND FLOW PRESSURE TRANSMITTER, the
technical specification is the document “FDSF6-114n-EDF NC-K3”.
In this example, the indication revision is the letter
n.
The drawings and schematics presented in this manual are not
contractual, so please refer to the specific dimensional plans of the
transmitter.
Please note on the electronic unit its configuration at the time of its
delivery before any manipulation that rotates it to avoid rotatating it
more than 90° and thus damaging the cable.
Important
Important