INITIAL STATE OF THE SETTING PARAMETERS
The electrical damping (time constant), the zero setting function (external screw) and that of the
measurement range, the transfer function, the display of the scale and the output signal in the
event of defect are set at the factory according to the table.
Each parameter can be changed with the FXW mobile communicator (HHC).
No.
Item
Settings
1
Damping
No damping
(Time constant)
(= periodic measurement)
2
Zero setting
Zero setting possible
with external screw
(Note1)
3
Transfer function
can be set to the factory according to
codification and the order
Digital display of the scale
Can be adjusted at the according to
codification and the order
4
(extraction of the square root)
7.07%
5
Output in the event of defect
linear
(square root setting)
6
Substitute values
Can be adjusted at the factory according
to codification and the order
7
Linearisation function
Not set