58
TN4FCXA2
j
-EDF-
E
K: Write protection (WRITE PROTCT)
The FCX-AII pressure transmitters
intended for use in nuclear power
plants are delivered configured
with an enabled write protection.
A password is required to unlock it.
The adjustment of the transmitter can be locked
(write protection) by entering a password
(PASSWORD)
If the lock is active, the write protection cannot
be deleted using the push buttons on the
digital indicator.
Refer to the menu “K: Write protection” for the
local setting using the push-buttons.
* The locking of the transmitter settings (write
protection) using the password is identical
to the locking function of the local setting
with the push buttons.
CL
K-1:
WRITE PROTCT
<
1
>
WRITE PR CHNG
<
2
>
PASSWORD SET
<
1
><
2
><
CL
>
3
ENT
ENT
2
K-2:
WRITE PROTCT
WRITE PROTECT
XXX
K-3:
WRITE PROTCT
PASSWORD
_
4
K-3:
WRITE PROTCT
PASSWORD
XXXXXXXX
CHNG OK?
<
ENT/CL
>
5
K-4:
WRITE WRITE
PASSWORD
XXXXXXXX
WRITE ERROR
<
CL
>
6
1
INC
K:
WRITE PROTCT
WRITE PROTECT
XXX
INC
If error of
PASSWORD
ENT
WRITE PROTECT
Possible to switch ON/OFF
<
INC
>
<
DEC
>
<
ENT
><
CL
>
<
INC
><
ENT
><
CL
>
CL
1
CL
INC
Menu No.L
Menu No.J
Menu No.L
<
INC
><
CHANGE
>
CHNG
CL
2
7
If PASSWORD
is correct
K:
WRITE PROTECT
WRITE PROTECT
XXX
<
INC
><
CHANGE
>
After keying
PASSWORD
1
CL
K-2:
WRITE PROTCT
OLD PASSWORD
_
8
ENT
2
7
K-3:
WRITE PROTCT
NEW PASSWORD
_
K-4:
WRITE PROTCT
NEW PASSWORD
XXXXXXXX
CHNG OK?
<
ENT/CL
>
9
K:
WRITE PROTCT
WRITE PROTECT
XXX
K-5:
WRITE WRITE
NEW PASSWORD
XXXXXXXX
WRITE ERROR
<
CL
>
1
10
CL
CL
<
ENT
><
CL
>
<
ENT
><
CL
>
<
INC
><
CHANGE
>
After keying
PASSWORD (old)
After keying
PASSWORD
(new)
ENT
CL
If error of
PASSWORD
If PASSWORD
is correct
Important