Chapter 4: Operation
4-44
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
CRRT treatments with citrate anticoagulation
4.6.7
Treatment
4.6.7.1
Checking the post-filter calcium concentration
The optical detector senses opaque fluid.
The blood pump is stopped.
If not done yet when using an NaCl
solution with two connectors:
Disconnect the venous patient line from
the NaCl solution and connect it to the
venous vascular access.
Use the rotary selector to select
Start
treatment? [OK] to confirm!
and press
[OK]
.
Note
Once treatment has started and the blood pump has achieved its target
delivery rate, the alarm limits are set automatically after approx.
10 seconds. Following this, the pressure alarm limits must be
checked by the operator and adjusted individually as required.
Warning
When starting the treatment, define the post-filter calcium value. If the
ionised calcium has not decreased at this point, it is absolutely
necessary that the tubing system and the solutions used are checked.
Warning
If the measured post-filter calcium value or systemic calcium value
varies significantly, please consult a physician. Observe the instructions
on taking a sample (see chapter 7.3.2 on page 7-17).
Summary of Contents for multiFiltrate
Page 1: ...multiFiltrate Instructions for Use Software version 5 3 Edition 15A 2015 Part no M56 963 1 ...
Page 2: ......
Page 12: ...xii Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 16: ...Chapter 1 Index 1 4 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 42: ...Chapter 3 Design 3 10 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 186: ...Chapter 5 Alarm processing 5 20 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 188: ...Chapter 6 Cleaning disinfection 6 2 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 218: ...Chapter 7 Functional description 7 30 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 228: ...Chapter 9 Installation 9 4 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 252: ...Chapter 12 Specifications 12 18 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 260: ...Chapter 13 Definitions 13 8 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 266: ...Chapter 14 Options 14 6 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 282: ...Chapter 15 Appendix 15 16 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...