Chapter 12: Specifications
12-6
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
Minimum distances
between radiated
interference source and
medical electrical devices
If a source of radiated interference uses a radio frequency technology
not specified in the table, the required minimum distance to medical
electrical devices can be obtained from the manufacturer.
RF transmission
technology
Minimum distance
from medical
electrical devices,
including all
connecting cables
Radiated
interference
source (examples)
(One device /
radiated interference
source may use
more than one RF
transmission
technology.)
GSM
(Global System for
Mobile
Communications)
3.3 m
Cell phone,
smartphone,
tablet computer
UMTS
(Universal Mobile
Telecommunications
System)
LTE
(Long-Term
Evolution)
DECT
(Digital Enhanced
Cordless
Telecommunications)
1.2 m
Cordless phone
WLAN
(Wireless Local Area
Network)
0.7 m
Notebook / laptop,
desktop computer,
e-book reader,
WLAN repeater,
WLAN router,
WLAN access point,
WLAN print server
Bluetooth
0.2 m
Wireless keyboard,
wireless mouse,
wireless
loudspeakers
Radio
0.1 m
Radio remote control
(except device-
specific radio remote
controls by the
manufacturer)
Summary of Contents for multiFiltrate
Page 1: ...multiFiltrate Instructions for Use Software version 5 3 Edition 15A 2015 Part no M56 963 1 ...
Page 2: ......
Page 12: ...xii Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 16: ...Chapter 1 Index 1 4 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 42: ...Chapter 3 Design 3 10 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 186: ...Chapter 5 Alarm processing 5 20 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 188: ...Chapter 6 Cleaning disinfection 6 2 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 218: ...Chapter 7 Functional description 7 30 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 228: ...Chapter 9 Installation 9 4 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 252: ...Chapter 12 Specifications 12 18 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 260: ...Chapter 13 Definitions 13 8 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 266: ...Chapter 14 Options 14 6 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 282: ...Chapter 15 Appendix 15 16 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...