Chapter 11: Technical Safety Checks / maintenance procedures
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
11-1
11 Technical Safety Checks / maintenance
procedures
11.1 Important information on the Technical Safety Checks /
maintenance procedures
Technical Safety Checks
(TSC)
The first TSC are required before the end of the 24th month following
initial start-up after delivery from the factory. All additional TSC are
required before the end of the 24th month following the last TSC
performed.
Maintenance procedures
(MA)
The maintenance procedures (MA) are a recommendation of the
manufacturer. The maintenance procedures help ensure trouble-free
operation, and must be carried out for the first time before the end of the
24th month following initial start-up after delivery from the factory. All
additional maintenance procedures are required before the end of the
24th month following the last maintenance procedure performed.
Qualification requirements
of testers
The checks must only be performed by the manufacturer's service
support organisation or a person authorised by it.
The checks may only be performed by personnel qualified to perform
them correctly based on their education, training, knowledge, and
experience. Furthermore, the personnel performing the checks must be
permitted to do so independently and without outside interference.
Specifications
Any information on the specifications must be observed.
Documentation
The Technical Safety Checks, the maintenance points, and further
procedural information are described in the Service Manual.
Reports are available on request.
The completion of the Technical Safety Checks must be entered in the
Medical Device Register.
Summary of Contents for multiFiltrate
Page 1: ...multiFiltrate Instructions for Use Software version 5 3 Edition 15A 2015 Part no M56 963 1 ...
Page 2: ......
Page 12: ...xii Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 16: ...Chapter 1 Index 1 4 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 42: ...Chapter 3 Design 3 10 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 186: ...Chapter 5 Alarm processing 5 20 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 188: ...Chapter 6 Cleaning disinfection 6 2 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 218: ...Chapter 7 Functional description 7 30 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 228: ...Chapter 9 Installation 9 4 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 252: ...Chapter 12 Specifications 12 18 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 260: ...Chapter 13 Definitions 13 8 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 266: ...Chapter 14 Options 14 6 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 282: ...Chapter 15 Appendix 15 16 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...