Chapter 7: Functional description
7-6
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
CVVHD
With CVVHD, the blood is purified mainly through dialysis. In addition to
solute transport through diffusion, which is the main process for
removing toxins in this case, a certain amount of convection also takes
place within the filter and by volume removal. Under typical CRRT
conditions, where the blood flow is considerably higher than the
dialysate flow, an almost complete saturation of the dialysate with toxins
of a low molecular weight, such as urea and creatinine, can normally be
expected. The efficacy of a CVVHD procedure is therefore comparable
with that of a Post CVVH procedure. As the speed of diffusion is relative
to the molecular weight, the full saturation of the dialysate with larger,
middle-molecular weight solutes may not be achieved, depending on
the blood and dialysate flow rates set. The clearance rate achieved for
these substances is thus lower than with Post CVVH (if the same
dialysate and substituate quantities are used). This disadvantage of
CVVHD can be compensated at least partially through the use of filters
with a large active surface and high-flux membranes. The advantage of
CVVHD lies in the possibility of setting a lower blood flow than in
Pre CVVH and Post CVVH.
Treatment parameters
CVVHD
(adults)
Min.
Max.
Resolution
Unit
Blood flow
10
500
10
ml/min
Ultrafiltration
Off / 10
1800
10
ml/h
UF goal
Off / 50
10,000
50
ml
Cont. heparin adm.
Off / 0.1
25
0.1
ml/h
Dialysate
600
4800
50
ml/h
Temperature
Off / 35
39
0.5
°C
Summary of Contents for multiFiltrate
Page 1: ...multiFiltrate Instructions for Use Software version 5 3 Edition 15A 2015 Part no M56 963 1 ...
Page 2: ......
Page 12: ...xii Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 16: ...Chapter 1 Index 1 4 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 42: ...Chapter 3 Design 3 10 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 186: ...Chapter 5 Alarm processing 5 20 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 188: ...Chapter 6 Cleaning disinfection 6 2 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 218: ...Chapter 7 Functional description 7 30 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 228: ...Chapter 9 Installation 9 4 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 252: ...Chapter 12 Specifications 12 18 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 260: ...Chapter 13 Definitions 13 8 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 266: ...Chapter 14 Options 14 6 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 282: ...Chapter 15 Appendix 15 16 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...