Chapter 4: Operation
4-2
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
Warning
Risk of cross-contamination if tubing systems without
hydrophobic filters are used
Only use tubing systems with hydrophobic filters in the pressure
lines.
Warning
Risk of cross-contamination in the event of an incorrect response
to a wet or defective hydrophobic filter
Never force back any fluid with a syringe
(risk of damaging the
hydrophobic filter)
.
Close off the tube with the wetted / defective hydrophobic filter.
Replace the relevant tubing system. In the case of a pressure line
with a wet hydrophobic filter, use a replacement pressure line
(accessory available from manufacturer).
If you cannot exclude the possibility that the device may have become
contaminated:
Take the device out of service after completing treatment.
Have the device tested for contamination by service support.
If the device is contaminated, all affected parts must be disinfected or
replaced by service support.
Warning
IV pole
Danger in case of excessive load (respect the maximum load)
If the load on the IV pole is too high, the device may tilt.
Ensure that the maximum allowed load on the IV pole of 6 kg is not
exceeded.
Warning
Tubing systems and multiFiltrate kits must be replaced no later than the
service life date indicated by the manufacturer. The usable life is
specified on the packaging of the tubing systems.
Summary of Contents for multiFiltrate
Page 1: ...multiFiltrate Instructions for Use Software version 5 3 Edition 15A 2015 Part no M56 963 1 ...
Page 2: ......
Page 12: ...xii Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 16: ...Chapter 1 Index 1 4 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 42: ...Chapter 3 Design 3 10 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 186: ...Chapter 5 Alarm processing 5 20 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 188: ...Chapter 6 Cleaning disinfection 6 2 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 218: ...Chapter 7 Functional description 7 30 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 228: ...Chapter 9 Installation 9 4 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 252: ...Chapter 12 Specifications 12 18 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 260: ...Chapter 13 Definitions 13 8 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 266: ...Chapter 14 Options 14 6 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 282: ...Chapter 15 Appendix 15 16 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...