Chapter 7: Functional description
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
7-15
7.2.5
Haemoperfusion (HP)
Adsorption
In haemoperfusion, toxins are removed by adsorption. It is employed to
remove toxic substances from the patient's blood which cannot be
removed by dialysis or haemofiltration due to proteopexy, for example.
It is a possible therapy following attempted suicide with certain drugs or
for the treatment of amanita poisoning. The extracorporeal blood circuit
is monitored as in conventional dialysis. Any possible clot formation in
the adsorber cartridge will be detected at an early stage by monitoring
the adsorber inlet pressure (pre-filter pressure) between the blood
pump and the adsorber cartridge.
Adsorber cartridge
The patient is connected as usual to an extracorporeal circuit that is
driven by a blood pump. Instead of flowing through a dialyser, the blood
passes through an adsorber cartridge filled, for example, with small
sorbent particles. The arrangement of the sorbent material serves to
increase the adsorption surface. Inlet and outlet port of the adsorber
cartridge are closed by sieves of appropriate mesh size which permit
the passage of blood cells but prevent the comparatively large sorbent
particles from escaping.
Treatment parameters
Available settings
Fig.: Schematic of a haemoperfusion treatment (HP)
Legend
1
Blood pump
2
Adsorber cartridge
3
Air detector
4
Venous clamp
HP
Min.
Max.
Resolution
Unit
Blood flow
10
500
10
ml/min
Cont. heparin adm.
Off / 0.1
25
0.1
ml/h
P
P
P
1
2
4
3
Summary of Contents for multiFiltrate
Page 1: ...multiFiltrate Instructions for Use Software version 5 3 Edition 15A 2015 Part no M56 963 1 ...
Page 2: ......
Page 12: ...xii Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 16: ...Chapter 1 Index 1 4 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 42: ...Chapter 3 Design 3 10 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 186: ...Chapter 5 Alarm processing 5 20 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 188: ...Chapter 6 Cleaning disinfection 6 2 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 218: ...Chapter 7 Functional description 7 30 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 228: ...Chapter 9 Installation 9 4 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 252: ...Chapter 12 Specifications 12 18 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 260: ...Chapter 13 Definitions 13 8 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 266: ...Chapter 14 Options 14 6 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 282: ...Chapter 15 Appendix 15 16 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...