Chapter 4: Operation
4-12
Fresenius Medical Care multiFiltrate IFU-EN-UK 15A-2015
4.4
Making the device ready for operation
4.4.1
Turning the device on
Use the rotary selector to select
Bolus
and press and hold
[OK]
.
As long as the selector is being pressed,
the bolus is slowly administered at a rate
of 0 ml up to 5 ml maximum (adjustable
via the System parameters menu). The
bolus will stop as soon as the selector is
released.
If the bolus, once activated, is stopped before
reaching the maximum volume, bolus
administration may be continued within
5 seconds until reaching the maximum
volume.
Note
Check the device for stability. The maximum angle of inclination must
not exceed 5°.
Before turning the device on, please ensure that all scales are
unloaded, that no objects are placed on the scales and that no tubing
system is inserted in the Ci-Ca pumps.
The maximum loading capacity of 12 kg per scale must not be
exceeded. The weighing cell can even be permanently damaged by a
short-term overload (e.g., pulling or lifting the device by the scales), in
which case the device can no longer be used.
Note
The operator must check all of the status indicator lights.
Once mains power is switched on, the green LED is illuminated
(dimmed). If the device is switched on via the "On / Off" key, the green
LED goes out and the yellow LED is illuminated. At the end of the
functional test, an audible alarm sounds and the red LED is illuminated
and then changes back to yellow.
Summary of Contents for multiFiltrate
Page 1: ...multiFiltrate Instructions for Use Software version 5 3 Edition 15A 2015 Part no M56 963 1 ...
Page 2: ......
Page 12: ...xii Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 16: ...Chapter 1 Index 1 4 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 42: ...Chapter 3 Design 3 10 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 186: ...Chapter 5 Alarm processing 5 20 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 188: ...Chapter 6 Cleaning disinfection 6 2 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 218: ...Chapter 7 Functional description 7 30 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 228: ...Chapter 9 Installation 9 4 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 252: ...Chapter 12 Specifications 12 18 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 260: ...Chapter 13 Definitions 13 8 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 266: ...Chapter 14 Options 14 6 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...
Page 282: ...Chapter 15 Appendix 15 16 Fresenius Medical Care multiFiltrate IFU EN UK 15A 2015 ...