The Piston-Cylinder selected must be able to cover the range of 75 kPa to
103 kPa absolute pressure.
PM200 Test Definition Matrices
Table 9. PM200-BG2.5K Pressure Point MatrixPM200-BG2.5K
As Found / As Left Points
Calibration Segment
Pressure No.
Set Pressure
[kPa]
Dwell Time [s]
Averaging Time
[s]
Ascending Points
1 -2.5
1
90
30
2 -1.5
90
30
3 -0.5
90
30
4 0.5
90
30
5 1.5
90
30
6 2.5
90
30
Descending Points
7 1.5
90
30
8 0.5
90
30
9 -0.5
90
30
10 -1.5
90 30
11 -2.5
90 30
1. UUT is zeroed before setting pressure.