Pressure
Controller/Calibrator
Maintenance
21
Clean the Exterior
To clean the Product, wipe it with a cloth that is lightly dampened with water or
mild detergent. To clean the display, use a soft cloth lightly dampened with
alcohol.
Caution
Do not use aromatic hydrocarbons or chlorinated solvents for
cleaning. They can damage the plastic materials used in the
Product.
Pressure Leakage Test (on-board diagnostics)
1. From the main menu, touch
Tasks
or
SETUP>Tasks
.
2. Touch
Leak Test
.
3. Complete these sections:
a. Setpoint – Pressure set point on which to perform the leak test.
b. Stability – After the set point is achieved, the time in seconds to wait for
stability before the duration of the test begins.
c. Duration – The time in seconds to collect data for leak-rate calculation.
d. Mode – Defaults to external, not selectable.
e. Run – Runs the test.
f. Exit – Exit menu.
Note
With the test port plugged and no volume, it is expected to achieve
better than 0.04 %FS/Min rate.
Summary of Contents for 6270A
Page 6: ...6270A Service Manual iv...
Page 26: ...6270A Service Manual 20 2 1 3 hxn212 eps Figure 7 PMM and PCM Removal...
Page 30: ...6270A Service Manual 24 1 2 3 hxn205 eps Figure 9 Manifold Installation...
Page 32: ...6270A Service Manual 26 2 1 1 hxn202 eps Figure 10 Remove Feet and Tilt Stands...
Page 40: ...6270A Service Manual 34 hxn208 eps Figure 16 Remove the CPU...
Page 43: ...Pressure Controller Calibrator Module Guide Rails 37 8 9 10 hxn219 eps Figure 18 Dock Assembly...
Page 64: ...6270A Service Manual 58 1 2 3 4 5 7 6 hxn210 eps Figure 32 Remove Latch Assembly...
Page 67: ...Pressure Controller Calibrator PCM Assembly 61 2 1 hxn401 eps Figure 34 Module Components...