Replace the Control Valves
Each valve and the blank have specific orientation which is referenced by a guide
pin in the manifold. Make sure the apply and release valves are not reversed.
When the valves are reversed, control performance is impacted and control
tuning is required. The control valve is the valve nearest to the front or in the
middle space if the blank is included, see Figure 35.
hxn403.eps
Figure 35. Replace the Control Valves (A)
To replace the control valve, see Figure 36:
1. Shut off supply pressure and vent the system.
2. Remove all pressure lines and fittings from the manifold.
3. Remove The PCM. See
PMM/PCM Removal
.
4. Remove the cover and clamp screws from the PCM. See
PCM O-rings,
Handle, Cover, and Clamp Screws
.
Note
To prevent leaks, only remove the blank or specific valve if an issue
is suspected.
5. Inspect or replace the blank only if it leaks.
a. Use a #2 Phillips screwdriver to remove the four M4 x 7 x 20 screws
.
b. Remove the blank and inspect the O-rings
.
c. Reverse these steps to install the blank. Use the guide pin
to correctly
orient the blank.
Summary of Contents for 6270A
Page 6: ...6270A Service Manual iv...
Page 26: ...6270A Service Manual 20 2 1 3 hxn212 eps Figure 7 PMM and PCM Removal...
Page 30: ...6270A Service Manual 24 1 2 3 hxn205 eps Figure 9 Manifold Installation...
Page 32: ...6270A Service Manual 26 2 1 1 hxn202 eps Figure 10 Remove Feet and Tilt Stands...
Page 40: ...6270A Service Manual 34 hxn208 eps Figure 16 Remove the CPU...
Page 43: ...Pressure Controller Calibrator Module Guide Rails 37 8 9 10 hxn219 eps Figure 18 Dock Assembly...
Page 64: ...6270A Service Manual 58 1 2 3 4 5 7 6 hxn210 eps Figure 32 Remove Latch Assembly...
Page 67: ...Pressure Controller Calibrator PCM Assembly 61 2 1 hxn401 eps Figure 34 Module Components...