Pressure
Controller/Calibrator
Calibration Tables
99
Table 26. PM200-G2.5M Pressure Point Matrix
As Found / As Left Points
Calibration Segment
Pressure No.
Set Pressure
[kPa]
Dwell Time
[s]
Averaging
Time [s]
Ascending Points
1 0
1
60
30
2 500
60
30
3 1000
60
30
4 1500
60
30
5 2000
60
30
6 2500
60
30
Descending Points
7 2000
60
30
8 1500
60
30
9 1000
60
30
10 500 60
30
11 0 60
30
1. UUT is zeroed before setting pressure.
Table 27. PM200-G3.5M Pressure Point Matrix
As Found / As Left Points
Calibration Segment
Pressure No.
Set Pressure
[kPa]
Dwell Time
[s]
Averaging
Time [s]
Ascending Points
1 0
1
60
30
2 700
60
30
3 1400
60
30
4 2100
60
30
5 2800
60
30
6 3500
60
30
Descending Points
7 2800
60
30
8 2100
60
30
9 1400
60
30
10 700 60
30
11 0 60
30
1. UUT is zeroed before setting pressure.
Summary of Contents for 6270A
Page 6: ...6270A Service Manual iv...
Page 26: ...6270A Service Manual 20 2 1 3 hxn212 eps Figure 7 PMM and PCM Removal...
Page 30: ...6270A Service Manual 24 1 2 3 hxn205 eps Figure 9 Manifold Installation...
Page 32: ...6270A Service Manual 26 2 1 1 hxn202 eps Figure 10 Remove Feet and Tilt Stands...
Page 40: ...6270A Service Manual 34 hxn208 eps Figure 16 Remove the CPU...
Page 43: ...Pressure Controller Calibrator Module Guide Rails 37 8 9 10 hxn219 eps Figure 18 Dock Assembly...
Page 64: ...6270A Service Manual 58 1 2 3 4 5 7 6 hxn210 eps Figure 32 Remove Latch Assembly...
Page 67: ...Pressure Controller Calibrator PCM Assembly 61 2 1 hxn401 eps Figure 34 Module Components...