Pressure
Controller/Calibrator
Module Guide Rails
39
G- and B-Series Valves
Use the procedure below to repair, replace, or clean the G- or B-series valves or
associated parts. The G- and B-series valves are listed in Table 4.
Note
Valves can be replaced without removing the entire dock assembly.
However, instructions for complete removal are included here.
1. Shut off the supply pressure and vent the system.
2. Remove all pressure lines and fittings from the manifold.
3. Use the procedures explained earlier to remove all PMMs and PCMs, top
panel, and COMM tray.
If the valves will be removed without completely removing the dock
assemblies, stop here. For complete removal of dock assemblies, also see
the sections
Manifold Replacement
and
PMM Dock Assembly
.
Valves come with coils and wires installed. The coils can be reused and only
replace the valve if the coil is still good. See Figure 20 and Table 3 for
wire/connector placement if installing new coil.
B4
B2
G3
B1
B1
B2
B3
B3
G1
G1
G2
G3
G2
hxn226.eps
Figure 20. G and B Series Valves
Summary of Contents for 6270A
Page 6: ...6270A Service Manual iv...
Page 26: ...6270A Service Manual 20 2 1 3 hxn212 eps Figure 7 PMM and PCM Removal...
Page 30: ...6270A Service Manual 24 1 2 3 hxn205 eps Figure 9 Manifold Installation...
Page 32: ...6270A Service Manual 26 2 1 1 hxn202 eps Figure 10 Remove Feet and Tilt Stands...
Page 40: ...6270A Service Manual 34 hxn208 eps Figure 16 Remove the CPU...
Page 43: ...Pressure Controller Calibrator Module Guide Rails 37 8 9 10 hxn219 eps Figure 18 Dock Assembly...
Page 64: ...6270A Service Manual 58 1 2 3 4 5 7 6 hxn210 eps Figure 32 Remove Latch Assembly...
Page 67: ...Pressure Controller Calibrator PCM Assembly 61 2 1 hxn401 eps Figure 34 Module Components...