6270A
Service Manual
86
5. Read zOffset (value always in psi), enter
UCOEF_SET[0]?
Run As Found data for pressures defined in the calibration matrices in
Calibration Tables
.
1. The As Found data determines the new C0 and C1 calibration coefficients
and calibration date.
2. The C0, C1, and zOffset values must be applied to the As Found data in
order to calculate the new C0 and C1 coefficients.
Directions to correct data for zOffset, C0 and C1:
Note
Do these calculations in an Excel spreadsheet and for each pressure
point.
1. Iteration_1 = (zOFF_Removed)
UUT pressure – zOffset (double check if zOffset is always in psi or unit of
measure).
2. Iteration_2 = (C0_Removed).
zOff_Removed – C0
3. Iteration_3 = (C1_Removed).
C0_Removed/C1
Use the calculated (C1_Removed) to find the new (New C1) with the formula
below:
1. New C1
New C1 =1/SLOPE(C1_Removed data, reference pressure data)
Note
Do this in an Excel spreadsheet where C1_Removed data =
known_y’s and reference pressure data = known_x’s.
Calculate the new C0. Using the New C1, you will apply it to the C1_Removed
data then calculate the Intercept (New C0).
1. New C1 Applied = (C1_Corrected).
C1_Removed*New C1
2. New C0.
INTERCEPT(C1_Corrected,reference pressure data)
Summary of Contents for 6270A
Page 6: ...6270A Service Manual iv...
Page 26: ...6270A Service Manual 20 2 1 3 hxn212 eps Figure 7 PMM and PCM Removal...
Page 30: ...6270A Service Manual 24 1 2 3 hxn205 eps Figure 9 Manifold Installation...
Page 32: ...6270A Service Manual 26 2 1 1 hxn202 eps Figure 10 Remove Feet and Tilt Stands...
Page 40: ...6270A Service Manual 34 hxn208 eps Figure 16 Remove the CPU...
Page 43: ...Pressure Controller Calibrator Module Guide Rails 37 8 9 10 hxn219 eps Figure 18 Dock Assembly...
Page 64: ...6270A Service Manual 58 1 2 3 4 5 7 6 hxn210 eps Figure 32 Remove Latch Assembly...
Page 67: ...Pressure Controller Calibrator PCM Assembly 61 2 1 hxn401 eps Figure 34 Module Components...