Pressure
Controller/Calibrator
Repair and Replacement
29
Top Panel
Warning
To prevent possible electrical shock, fire or personal injury,
turn off the Product and disconnect the mains power cord prior
to the removal procedure. Shut off the supply pressure and
vent the system before you remove the manifold, valve, or
associated parts.
To remove the top panel, see Figure 12:
1. Make sure the Product is off and disconnect the mains power cable.
2. Remove the six M4 x 6 flat-heat screws with a #2 Phillips head
screwdriver.
3. Lift and remove the top panel.
hxn204.eps
Figure 12. Remove Top Panel