Pressure
Controller/Calibrator
Repair and Replacement
25
Repair and Replacement
The subsequent procedures explain how to disassemble parts of the Product. To
reinstall the parts, reverse the steps unless otherwise specified.
Feet and Tilt Stands
To remove or replace the feet and tilt stands, see Figure 10:
1. Shut off the supply pressure and vent the system.
2. Open the door to vent off the pressure between PCM and supply isolation
valve.
3. Make sure the Product is off and disconnect the mains power cord.
4. Remove all pressure lines and fittings from the manifold.
5. Turn over the chassis so the feet
and tilt stands
are face up.
6. Remove the feet and tilt stands with a 3/16 inch flat-head screw driver.
7. Attach tilt stands to the feet as shown.
8. Install two feet and two tilt stands on the front of the chassis.
9. Install two feet on the rear of chassis.
Summary of Contents for 6270A
Page 6: ...6270A Service Manual iv...
Page 26: ...6270A Service Manual 20 2 1 3 hxn212 eps Figure 7 PMM and PCM Removal...
Page 30: ...6270A Service Manual 24 1 2 3 hxn205 eps Figure 9 Manifold Installation...
Page 32: ...6270A Service Manual 26 2 1 1 hxn202 eps Figure 10 Remove Feet and Tilt Stands...
Page 40: ...6270A Service Manual 34 hxn208 eps Figure 16 Remove the CPU...
Page 43: ...Pressure Controller Calibrator Module Guide Rails 37 8 9 10 hxn219 eps Figure 18 Dock Assembly...
Page 64: ...6270A Service Manual 58 1 2 3 4 5 7 6 hxn210 eps Figure 32 Remove Latch Assembly...
Page 67: ...Pressure Controller Calibrator PCM Assembly 61 2 1 hxn401 eps Figure 34 Module Components...