6270A
Service Manual
68
PMM
PMM
PMM
PMM
V
e
nt
Reference
Te
s
t
Exhaust
Supply
PMM
PCM
6270A
High Range Module ≥ 1500 psi
Ball Valve
Regulator
Max 6000 psi Input
3300 psi Out
Vac
Pump
3300
6000
>3000 psi
N2 Bottle or Booster
Supply Bottle
CGA Fitting
+
hxn301.eps
Figure 40. Calibrating Sensors
Supply Port (ATM and 3,000 psi)
Note
There is an isolation valve between the supply port on the manifold
and the supply port of the PCM. Follow directions to ensure that the
supply pressure and valve position is correct for the calibration point.
Send the
CAL:PRES6:VAL1 xx.xx
command to open the supply
isolation valve. This command will be used twice.
1. Remove all supply pressure from the supply port of the Product. See
Figure 41.
2. Use a TEE
to connect an external digital pressure device that reads in
Absolute units
to the supply line. Use this as the reference pressure.
3. Touch
Vent
and then
Enter
to vent the Product.
4. Ensure the supply line is disconnected from the supply port.
5. With the reference and supply sensor vented to atmosphere, type
CAL:PRES6:VAL1 xx.xx
(xx.xx =value in units must be absolute). This
Summary of Contents for 6270A
Page 6: ...6270A Service Manual iv...
Page 26: ...6270A Service Manual 20 2 1 3 hxn212 eps Figure 7 PMM and PCM Removal...
Page 30: ...6270A Service Manual 24 1 2 3 hxn205 eps Figure 9 Manifold Installation...
Page 32: ...6270A Service Manual 26 2 1 1 hxn202 eps Figure 10 Remove Feet and Tilt Stands...
Page 40: ...6270A Service Manual 34 hxn208 eps Figure 16 Remove the CPU...
Page 43: ...Pressure Controller Calibrator Module Guide Rails 37 8 9 10 hxn219 eps Figure 18 Dock Assembly...
Page 64: ...6270A Service Manual 58 1 2 3 4 5 7 6 hxn210 eps Figure 32 Remove Latch Assembly...
Page 67: ...Pressure Controller Calibrator PCM Assembly 61 2 1 hxn401 eps Figure 34 Module Components...