6270A
Service Manual
108
Table 44. PM600- A10M Pressure Point Matrix
As Found / As Left Points
Calibration Segment
Pressure No.
Set Pressure
[kPa]
Dwell Time [s]
Averaging Time
[s]
Ascending Points
1 ATM
1
60 30
2 2500
60 30
3 5000
60 30
4 7500
60 30
5 10000
60 30
Descending Points
6 7500
60 30
7 5000
60 30
8 2500
60 30
9 ATM
1
60 30
1. ATM: atmospheric pressure from laboratory barometer via PG system.
Table 45. PM600- A14M Pressure Point Matrix
As Found / As Left Points
Calibration Segment
Pressure No.
Set Pressure
[kPa]
Dwell Time [s]
Averaging
Time [s]
Ascending Points
1 ATM
1
60 30
2 3500
60 30
3 7000
60 30
4 10500
60 30
5 14000
60 30
Descending Points
6 10500
60 30
7 7000
60 30
8 3500
60 30
9 ATM
1
60 30
1. ATM: atmospheric pressure from laboratory barometer via PG system.
Summary of Contents for 6270A
Page 6: ...6270A Service Manual iv...
Page 26: ...6270A Service Manual 20 2 1 3 hxn212 eps Figure 7 PMM and PCM Removal...
Page 30: ...6270A Service Manual 24 1 2 3 hxn205 eps Figure 9 Manifold Installation...
Page 32: ...6270A Service Manual 26 2 1 1 hxn202 eps Figure 10 Remove Feet and Tilt Stands...
Page 40: ...6270A Service Manual 34 hxn208 eps Figure 16 Remove the CPU...
Page 43: ...Pressure Controller Calibrator Module Guide Rails 37 8 9 10 hxn219 eps Figure 18 Dock Assembly...
Page 64: ...6270A Service Manual 58 1 2 3 4 5 7 6 hxn210 eps Figure 32 Remove Latch Assembly...
Page 67: ...Pressure Controller Calibrator PCM Assembly 61 2 1 hxn401 eps Figure 34 Module Components...