6270A
Service Manual
112
Table 48. Troubleshooting (cont.)
Problem Probable
Cause
Action
Pressure Generation or Indication Problems
Product will not enter
control mode
Modules not
installed (PCM and
1+ modules)
Verify modules are properly installed. Door must be
closed and properly sensed as closed. Confirm door
status and pressure supply to PCM.
PMMs not fully
installed
Follow steps outlined in
Installation and Setup
to ensure
that all modules are properly installed.
No pressure
supply
Apply proper supply pressure.
Door not closed
Make sure the front panel is closed.
Internal door-close
sensor failed
Check the sensor. Send the Product to Fluke Calibration
for repair.
System in remote
operation mode
No pressure
measurement is
shown
Door not closed
Verify front panel is closed, verify the USB connection to
front panel is proper, verify Product is installed
PMM not installed
Install a PMM (see
Installation and Setup)
PCM not installed
Install a PCM (see
Installation and Setup)
Product does not
reach target pressure
Pressure supply
too low
Verify pressure supply, perform leak check.
Leak
Verify Test port is leak tight.
Valves require
service
Send the Product to Fluke Calibration for repair.
Target pressure
set higher than the
highest ranged
PMM.
Install a PMM with the appropriate range.
Slew Rate set to
zero
Increase slew rate.
While in measure
mode, pressure leaks
No supply
pressure
Provide supply pressure if not connected.
Summary of Contents for 6270A
Page 6: ...6270A Service Manual iv...
Page 26: ...6270A Service Manual 20 2 1 3 hxn212 eps Figure 7 PMM and PCM Removal...
Page 30: ...6270A Service Manual 24 1 2 3 hxn205 eps Figure 9 Manifold Installation...
Page 32: ...6270A Service Manual 26 2 1 1 hxn202 eps Figure 10 Remove Feet and Tilt Stands...
Page 40: ...6270A Service Manual 34 hxn208 eps Figure 16 Remove the CPU...
Page 43: ...Pressure Controller Calibrator Module Guide Rails 37 8 9 10 hxn219 eps Figure 18 Dock Assembly...
Page 64: ...6270A Service Manual 58 1 2 3 4 5 7 6 hxn210 eps Figure 32 Remove Latch Assembly...
Page 67: ...Pressure Controller Calibrator PCM Assembly 61 2 1 hxn401 eps Figure 34 Module Components...