Pressure
Controller/Calibrator
Calibration Tables
97
Table 22. PM200-G700K Pressure Point Matrix
As Found / As Left Points
Calibration Segment
Pressure No.
Set Pressure
[kPa]
Dwell Time [s]
Averaging Time
[s]
Ascending Points
1 0
1
60 30
2 140
60 30
3 280
60 30
4 420
60 30
5 560
60 30
6 700
60 30
Descending Points
7 560
60 30
8 420
60 30
9 280
60 30
10 140 60 30
11 0 60 30
1. UUT is zeroed before setting pressure.
Table 23. PM200-G1M Pressure Point Matrix
As Found / As Left Points
Calibration Segment
Pressure No.
Set Pressure
[kPa]
Dwell Time [s]
Averaging Time [s]
Ascending Points
1 0
1
60 30
2 200
60
30
3 400
60
30
4 600
60
30
5 800
60
30
6 1000
60
30
Descending Points
7 800
60
30
8 600
60
30
9 400
60
30
10 200 60 30
11 0 60 30
1. UUT is zeroed before setting pressure.
Summary of Contents for 6270A
Page 6: ...6270A Service Manual iv...
Page 26: ...6270A Service Manual 20 2 1 3 hxn212 eps Figure 7 PMM and PCM Removal...
Page 30: ...6270A Service Manual 24 1 2 3 hxn205 eps Figure 9 Manifold Installation...
Page 32: ...6270A Service Manual 26 2 1 1 hxn202 eps Figure 10 Remove Feet and Tilt Stands...
Page 40: ...6270A Service Manual 34 hxn208 eps Figure 16 Remove the CPU...
Page 43: ...Pressure Controller Calibrator Module Guide Rails 37 8 9 10 hxn219 eps Figure 18 Dock Assembly...
Page 64: ...6270A Service Manual 58 1 2 3 4 5 7 6 hxn210 eps Figure 32 Remove Latch Assembly...
Page 67: ...Pressure Controller Calibrator PCM Assembly 61 2 1 hxn401 eps Figure 34 Module Components...