Pressure
Controller/Calibrator
Calibration Tables
109
Table 46. PM600- A20M Pressure Point Matrix
As Found / As Left Points
Calibration Segment
Pressure No.
Set Pressure
[kPa]
Dwell Time [s]
Averaging
Time [s]
Ascending Points
1 ATM
1
60 30
2 5000
60 30
3 10
000
60 30
4 15
000
60 30
5 20
000
60 30
Descending Points
6 15
000
60 30
7 10
000
60 30
8 5000
60 30
9 ATM
1
60 30
1. ATM: atmospheric pressure from laboratory barometer via PG system.
Table 47. BRM600-BA100K Pressure Point Matrix
As Found / As Left Points
Calibration Segment
Pressure No.
Set Pressure
[kPa]
Dwell Time [s]
Averaging Time
[s]
Ascending Points
1 70
60 30
2 80
60 30
3 90
60 30
4 100
60 30
5 110
60 30
Descending Points
6 100
60 30
7 90
60 30
8 80
60 30
9 70
60 30