Pressure
Controller/Calibrator
Software Upgrades
17
System Boot
If the system only partially boots up the screen on the left is shown. Check that
the PCM is properly installed. The screen on the right indicates that the system
has loaded and the application starts.
hxn410.jpg
Software Upgrades
To upgrade the software for the mainframe and interface:
1. Go to the Fluke Calibration website and download the latest 6270A firmware.
2. Connect a PC directly to the rear panel USB port.
Note
When the Product is connected to a PC with the USB connection,
the Product appears as a flash drive, usually assigned to the E:
drive. The drive letter varies depending upon the PC setup.
3. Copy the file:
script6270
and either/both
6270A_main
or
6270A_gui
to
the flash drive.
4. Wait 30 seconds.
5. Turn off and then turn on the Product.
The firmware upgrades at system bootup.
To upgrade the software for the PCM:
1. From the Fluke Calibration website, download the latest PCM software to
your PC and rename it
6270A_pcm.img
.
2. Connect a PC directly to the rear panel USB port.
3. Copy
6270A_pcm.img
and
Script6270
to the E: drive.
4. Wait 30 seconds.
5. Turn off and then turn on the Product.
6. The process is complete when the
Verifying PCM Installation…
screen
disappears.
Summary of Contents for 6270A
Page 6: ...6270A Service Manual iv...
Page 26: ...6270A Service Manual 20 2 1 3 hxn212 eps Figure 7 PMM and PCM Removal...
Page 30: ...6270A Service Manual 24 1 2 3 hxn205 eps Figure 9 Manifold Installation...
Page 32: ...6270A Service Manual 26 2 1 1 hxn202 eps Figure 10 Remove Feet and Tilt Stands...
Page 40: ...6270A Service Manual 34 hxn208 eps Figure 16 Remove the CPU...
Page 43: ...Pressure Controller Calibrator Module Guide Rails 37 8 9 10 hxn219 eps Figure 18 Dock Assembly...
Page 64: ...6270A Service Manual 58 1 2 3 4 5 7 6 hxn210 eps Figure 32 Remove Latch Assembly...
Page 67: ...Pressure Controller Calibrator PCM Assembly 61 2 1 hxn401 eps Figure 34 Module Components...