FLUXUS F709: Connection to the transmitter without cable gland
For the assignment and the activation of the temperature inputs see chapter 17.
7.2.1
Direct Connection of the Temperature Probe
FLUXUS F704
• Remove the second blind plug on the right for the connection of the temperature probe (see Fig. 7.25).
• Open the cable gland of the temperature probe. The compression part remains in the cap nut (see Fig. 7.26 and Tab.
• Push the cable of the temperature probe through the cap nut, the compression part, the basic part, and the reducer.
• Prepare the cable.
• Insert the cable into the housing.
• Screw the gasket ring side of the reducer tightly into the housing of the transmitter.
• Tightly screw the basic part into the reducer.
• Fix the cable gland by screwing the cap nut onto the basic part.
• Connect the temperature probe to the terminals of the transmitter (see Fig. 7.1, Fig. 7.25 and Tab. 7.13).
Tab. 7.12: Connection systems
direct connection
connection with
extension cable
connection with
junction box
Attention!
Observe the "Safety Instructions for the Use in Explosive Atmosphere" (see document SIFLUXUS).
te
mper
atu
re prob
e
transmitter
1
ext
ension
cab
le
te
mpera
tur
e probe
transmitter
transmitter
1
1
ex
te
ns
io
n
c
ab
le
te
mper
atu
re prob
e
transmitter
1
Summary of Contents for FLUXUS F704**-A2 Series
Page 4: ...FLUXUS F70x UMFLUXUS_F7V4 6 2EN 2017 10 01...
Page 8: ...FLUXUS F70x Table of Contents 8 UMFLUXUS_F7V4 6 2EN 2017 10 01...
Page 191: ......
Page 192: ......