6 Installation
48
UMFLUXUS_F7V4-6-2EN, 2017-10-01
6.5.2.3
Installation with a Quick Release Clasp
• Cut the tension strap to length (pipe circumf at least 120 mm).
• Position the temperature probe on the pipe (see Fig. 6.58).
• Place the tension strap around the temperature probe and the pipe.
• Insert the tension strap into the clasp (see Fig. 6.60).
• Tighten the tension strap.
• Tighten the screw of the clasp.
6.5.3
Installation of the Temperature Probe (Response Time 8 s)
• Fix the protection plate and the insulation foam to the temperature probe (see Fig. 6.61).
• Take the spring end of the ball chain and insert the last ball into one of the slots on the upper side of the temperature
probe (see Fig. 6.62).
• Place the chain around the pipe. Tighten the chain and insert it into the other slot of the temperature probe.
Fig. 6.60: Quick release clasp
Fig. 6.61: Temperature probe
Note!
The entire contact surface of the temperature probe always has to rest on the pipe. In case of very
small pipes, the protection plate and the insulation foam have to be cut to size, if necessary.
Fig. 6.62: Temperature probe on the pipe
temperature Probe
protection plate
insulation foam
spring end
chain
slots on the upper
side of the
temperature probe