18 Outputs
UMFLUXUS_F7V4-6-2EN, 2017-10-01
161
18.8
Deactivation of the Outputs
If the programmed outputs are no longer required, they can be deactivated. The configuration of a deactivated output is
stored and will be available if the output is activated again.
Tab. 18.6: Pictograms for the alarm state indication
Nr.
func
(switching condition)
typ
(holding behavior)
mode
(switching function)
current state
R
=
1
OFF
NON-HOLD
NO Cont.
closed
2
MAX
HOLD
NC Cont.
open
3
MIN
+
→
- -
→
+
QUANT.
ERROR
Example:
R1 =
Select
no
in
Output Options\Alarm Output
to deactivate an output. Press ENTER.
Alarm Output
>NO<
yes